Title: Hybrid time–frequency domain dual-probe coherent anti-Stokes Raman scattering for simultaneous temperature and pressure measurements in compressible flows via spectral fitting

Journal Article · · Applied Optics
DOI:https://doi.org/10.1364/AO.472831· OSTI ID:1905177

We demonstrate a hybrid time–frequency spectroscopic method for simultaneous temperature/pressure measurements in nonreacting compressible flows with known gas composition. Hybrid femtosecond–picosecond, pure-rotational coherent anti-Stokes Raman scattering (CARS), with two independent, time-delayed probe pulses, is deployed for single-laser-shot measurements of temperature and pressure profiles along an ∼ <#comment/> 5 -mm line. The theory of dual-probe CARS is presented, along with a discussion of the iterative fitting of experimental spectra. Temperature is obtained from spectra acquired with an early, near-collision-free probe time delay ( τ <#comment/> 1 = 0 p s ) and pressure from spectra obtained at probe delays of τ <#comment/> 2 = 150 − <#comment/> 1000 p s , where collisions significantly impact the spectral profile. Unique solutions for temperature and pressure are obtained by iteratively fitting the two spectra to account for small collisional effects observed for the near zero probe delay spectrum. A dual-probe pure-rotational CARS system, in a 1D line-imaging configuration, is developed to demonstrate effectively the simultaneous temperature and pressure profiles recorded along the axial centerline of a highly underexpanded jet. The underexpanded air jet permits evaluation of this hybrid time–frequency domain approach for temperature and pressure measurements across a wide range of low-temperature–low-pressure conditions of interest in supersonic ground-test facilities. Single-laser-shot measurement precisions in both quantities and pressure measurement accuracy are systematically evaluated in the quiet zone upstream of the Mach disk. Precise thermometry approaching 1%−2% is observed in regions of high CARS signal-to-noise ratios. Pressure measurements are optimized at probe time delays where the ratio of the late probe delay to the Raman lifetime exceeds four ( τ <#comment/> 2 / τ <#comment/> R > <#comment/> 4 ). The impact of low-temperature Raman linewidths on CARS pressure measurements is evaluated, and comparisons of CARS pressures obtained with our recent low-temperature pure-rotational Raman linewidth data and extrapolated high-temperature Q -branch linewidths are presented. Considering all measurements with τ <#comment/> 2 / τ <#comment/> R ≥ <#comment/> 4.0 , measured pressures were on average 7.9% of the computed isentropic values with average shot-to-shot deviations representing a combination of instrument noise and fluid fluctuations of  5.0%.

Sponsoring Organization:
USDOE National Nuclear Security Administration (NNSA)
Grant/Contract Number:
NA0002374; NA0003525; NA0003963
OSTI ID:
1905177
Journal Information:
Applied Optics, Journal Name: Applied Optics Vol. 62 Journal Issue: 1; ISSN 1559-128X
Publisher:
Optical Society of AmericaCopyright Statement
Country of Publication:
United States
Language:
English

References (57)

Simultaneous measurements of velocity, temperature, and pressure using rapid cw wavelength-modulation laser-induced fluorescence of OH journal June 1990
Diode-laser absorption technique for simultaneous measurements of multiple gasdynamic parameters in high-speed flows containing water vapor journal May 1994
Multiparameter spatio-thermochemical probing of flame–wall interactions advanced with coherent Raman imaging journal January 2017
On the sensitivity of rotational CARS N2 thermometry to the Herman-Wallis factor journal February 2011
Free underexpanded jets in a quiescent medium: A review journal August 2015
High resolution rotational CARS spectrum of oxygen journal February 1992
Recent advances in coherent anti-Stokes Raman scattering spectroscopy: Fundamental developments and applications in reacting flows journal April 2010
Dual-Pump Coherent Anti-Stokes Raman Scattering for Simultaneous Pressure/Temperature Measurement journal February 1998
Effects of imaging system blur on measurements of flow scalars and scalar gradients journal May 2004
Determination of N 2 –N 2 and N 2 –O 2 S-branch Raman linewidths using time-resolved picosecond pure rotational coherent anti-Stokes Raman scattering journal January 2019
Rotational Raman intensity-correction factors due to vibrational anharmonicity: their effect on temperature measurements journal January 1982
Nonintrusive Pressure and Temperature Measurements in an Underexpanded Sonic Jet Flowfield journal June 2004
Experimental and theoretical analysis of the temperature dependence of rotational Raman linewidths of oxygen journal January 1983
Communication: Time-domain measurement of high-pressure N2 and O2 self-broadened linewidths using hybrid femtosecond/picosecond coherent anti-Stokes Raman scattering journal November 2011
On the use of filtered Rayleigh scattering for measurements in compressible flows and thermal fields journal December 2004
Pressure, Temperature and Velocity Measurements in Underexpanded Free Jets using Laser-Induced Fluorescence Imaging journal April 2009
Split-probe hybrid femtosecond/picosecond rotational CARS for time-domain measurement of S-branch Raman linewidths within a single laser shot journal January 2013
Time-domain measurements of S-branch N2–N2 Raman linewidths using picosecond pure rotational coherent anti-Stokes Raman spectroscopy journal May 2012
Pure-rotational H2 thermometry by ultrabroadband coherent anti-Stokes Raman spectroscopy journal June 2017
Development of two-beam femtosecond/picosecond one-dimensional rotational coherent anti-Stokes Raman spectroscopy: Time-resolved probing of flame wall interactions journal January 2015
Gas-phase temperature measurement in the vaporizing spray of a gasoline direct-injection injector by use of pure rotational coherent anti-Stokes Raman scattering journal January 2004
A test of different rotational Raman linewidth models: Accuracy of rotational coherent anti‐Stokes Raman scattering thermometry in nitrogen from 295 to 1850 K journal August 1993
Picosecond time-resolved pure-rotational coherent anti-Stokes Raman spectroscopy for N_2 thermometry journal January 2009
Communication: Simplified two-beam rotational CARS signal generation demonstrated in 1D journal February 2013
Interference-free gas-phase thermometry at elevated pressure using hybrid femtosecond/picosecond rotational coherent anti-Stokes Raman scattering journal January 2012
Development of high-resolution N_2 coherent anti-Stokes Raman scattering for measuring pressure, temperature, and density in high-speed gas flows journal January 2000
Studies of nitrogen self-broadening at high temperature with inverse Raman spectroscopy journal January 1986
Subaperture sampling for digital-holography applications involving atmospheric turbulence journal June 2021
Hybrid fs/ps rotational CARS temperature and oxygen measurements in the product gases of canonical flat flames journal May 2015
Oxygen rotational Raman linewidth determination considering nonmonoexponential decoherence behavior journal October 2018
Dielectric-barrier-discharge plasma-assisted hydrogen diffusion flame. Part 1: Temperature, oxygen, and fuel measurements by one-dimensional fs/ps rotational CARS imaging journal May 2018
Dual-probe 1D hybrid fs/ps rotational CARS for simultaneous single-shot temperature, pressure, and O 2 /N 2 measurements journal January 2020
Hybrid femtosecond/picosecond coherent anti-Stokes Raman scattering for high-speed gas-phase thermometry journal January 2010
Development of a combined pure rotational and vibrational coherent anti-Stokes Raman scattering system journal January 2013
Rotational coherence beating in molecular oxygen: Coupling between electronic spin and nuclear angular momenta journal December 2018
Single-shot ultrafast coherent anti-Stokes Raman scattering of vibrational/rotational nonequilibrium journal January 2017
Demonstration and characterization of filtered Rayleigh scattering for planar velocity measurements journal March 1996
Low frequency coherent anti-Stokes Raman spectroscopy of air journal November 1978
High-Resolution Broadband N2 Coherent Anti-Stokes Raman Spectroscopy: Comparison of Measurements for Conventional And Modeless Broadband Dye Lasers journal January 2003
High performance resonance Raman spectroscopy using volume Bragg gratings as tunable light filters journal May 2010
Multi-Property Measurements at High Sampling Rates Using Rayleigh Scattering journal April 2009
Volume Bragg gratings as ultra-narrow and multiband optical filters conference June 2012
Calculation of collisionally narrowed coherent anti-Stokes Raman spectroscopy spectra journal October 1985
Continuous wave dye-laser technique for simultaneous, spatially resolved measurements of temperature, pressure, and velocity of NO in an underexpanded free jet journal July 1993
Temperature and density dependence of the linewidths and line shifts of the rotational Raman lines in N 2 and H 2 journal September 1986
High-resolution CARS measurements of temperature profiles and pressure in a tungsten lamp journal January 1987
Time-domain self-broadened and air-broadened nitrogen S-branch Raman linewidths at 80–200 K recorded in an underexpanded jet journal May 2022
Rotational temperature and density measurements in a hypersonic flow by dual-line CARS journal January 1993
Pressure measurements using hybrid femtosecond/picosecond rotational coherent anti-Stokes Raman scattering journal January 2015
Laser diode wavelength-modulation spectroscopy for simultaneous measurement of temperature, pressure, and velocity in shock-heated oxygen flows journal January 1993
Collisional Raman linewidths of nitrogen at high temperature (1700–2400 K) journal May 1995
Simulation des Intensitätsverlaufs im Raman-Spektrum von Sauerstoff unter Berücksichtigung der Spinaufspaltung journal January 1972
Pressure measurement in combusting and non-combusting gases using laser-induced grating spectroscopy journal February 2019
Temperature measurements in metalized propellant combustion using hybrid fs/ps coherent anti-Stokes Raman scattering journal January 2016
Gas-phase pressure measurement using femtosecond laser-induced grating scattering technique journal March 2022
Hybrid femtosecond/picosecond rotational coherent anti-Stokes Raman scattering temperature and concentration measurements using two different picosecond-duration probes journal January 2013
Comparison of chirped-probe-pulse and hybrid femtosecond/picosecond coherent anti-Stokes Raman scattering for combustion thermometry journal January 2017