Transfer of autocollimator calibration for use with scanning gantry profilometers for accurate determination of surface slope and curvature of state-of-the-art x-ray mirrors

Journal Article · · Advances in Metrology for X-Ray and EUV Optics VIII
DOI:https://doi.org/10.1117/12.2529519· OSTI ID:1887190
 [1];  [1];  [1];  [2];  [2];  [3];  [1];  [1]
  1. Lawrence Berkeley National Lab. (LBNL), Berkeley, CA (United States)
  2. Physikalisch-Technische Bundesanstalt (PTB), Braunschweig (Germany)
  3. Helmholtz-Zentrum Berlin (HZB), (Germany). German Research Centre for Materials and Energy
X-ray optics, desired for beamlines at free-electron-laser and diffraction-limited-storage-ring x-ray light sources, must have almost perfect surfaces, capable of delivering light to experiments without significant degradation of brightness and coherence. To accurately characterize such optics at an optical metrology lab, two basic types of surface slope profilometers are used: The long trace profilers (LTPs) and nanometer optical measuring (NOM) like angular deflectometers, based on electronic autocollimator (AC) ELCOMAT-3000. The inherent systematic errors of the instrument's optical sensors set the principle limit to their measuring performance. Where autocollimator of a NOM-like profiler may be calibrated at a unique dedicated facility, this is for a particular configuration of distance, aperture size, and angular range that does not always match the exact use in a scanning measurement with the profiler. Here we discuss the developed methodology, experimental set-up, and numerical methods of transferring the calibration of one reference AC to the scanning AC of the Optical Surface Measuring System (OSMS), recently brought to operation at the ALS Xray Optics Laboratory. We show that precision calibration of the OSMS performed in three steps, allows us to provide high confidence and accuracy low-spatial-frequency metrology and not â€print into' measurements the inherent systematic error of tool in use. Here, with the examples of the OSMS measurements with a state-of-The-Art x-ray aspherical mirror, available from one of the most advanced vendors of X-ray optics, we demonstrate the high efficacy of the developed calibration procedure. The results of our work are important for obtaining high reliability data, needed for sophisticated numerical simulations of beamline performance and optimization of beamline usage of the optics. This work was supported by the U. S. Department of Energy under contract number DE-AC02-05CH11231.
Research Organization:
Lawrence Berkeley National Laboratory (LBNL), Berkeley, CA (United States)
Sponsoring Organization:
USDOE Office of Science (SC), Basic Energy Sciences (BES)
Grant/Contract Number:
AC02-05CH11231
OSTI ID:
1887190
Journal Information:
Advances in Metrology for X-Ray and EUV Optics VIII, Journal Name: Advances in Metrology for X-Ray and EUV Optics VIII Vol. 11109
Publisher:
SPIECopyright Statement
Country of Publication:
United States
Language:
English

References (61)

Improved measurement accuracy in a long trace profiler: Compensation for laser pointing instability journal August 1994
The Diamond-NOM: A non-contact profiler capable of characterizing optical figure error with sub-nanometre repeatability
  • Alcock, S. G.; Sawhney, K. J. S.; Scott, S.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 616, Issue 2-3 https://doi.org/10.1016/j.nima.2009.10.137
journal May 2010
Sub-microradian surface slope metrology with the ALS Developmental Long Trace Profiler
  • Yashchuk, Valeriy V.; Barber, Samuel; Domning, Edward E.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 616, Issue 2-3 https://doi.org/10.1016/j.nima.2009.10.175
journal May 2010
Autocollimators for deflectometry: Current status and future progress
  • Geckeler, Ralf D.; Just, Andreas; Krause, Michael
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 616, Issue 2-3 https://doi.org/10.1016/j.nima.2009.11.021
journal May 2010
Upgrade of long trace profiler for characterization of high-precision X-ray mirrors at SPring-8
  • Senba, Y.; Kishimoto, H.; Ohashi, H.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 616, Issue 2-3 https://doi.org/10.1016/j.nima.2009.12.007
journal May 2010
Characterization and calibration of 2nd generation slope measuring profiler
  • Siewert, Frank; Buchheim, Jana; Zeschke, Thomas
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 616, Issue 2-3 https://doi.org/10.1016/j.nima.2009.12.033
journal May 2010
Performance of the APS optical slope measuring system
  • Qian, Jun; Sullivan, Joe; Erdmann, Mark
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 710 https://doi.org/10.1016/j.nima.2012.10.102
journal May 2013
Characterization of the error budget of Alba-NOM
  • Nicolas, Josep; Martínez, Juan Carlos
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 710 https://doi.org/10.1016/j.nima.2012.10.125
journal May 2013
Investigations on the spatial resolution of autocollimator-based slope measuring profilers
  • Siewert, F.; Buchheim, J.; Höft, T.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 710 https://doi.org/10.1016/j.nima.2012.10.130
journal May 2013
Development of a high-performance gantry system for a new generation of optical slope measuring profilers
  • Assoufid, Lahsen; Brown, Nathan; Crews, Dan
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 710 https://doi.org/10.1016/j.nima.2012.11.063
journal May 2013
Approaching sub-50 nanoradian measurements by reducing the saw-tooth deviation of the autocollimator in the Nano-Optic-Measuring Machine
  • Qian, Shinan; Geckeler, Ralf D.; Just, Andreas
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 785 https://doi.org/10.1016/j.nima.2015.02.065
journal June 2015
Comparison of angle standards with the aid of a high-resolution angle encoder journal October 2009
Using a straightness reference in obtaining more accurate surface profiles from a long trace profiler journal January 1992
Portable long trace profiler: Concept and solution journal August 2001
Technique for measuring the groove density of diffraction gratings using the long trace profiler journal July 2003
The Nanometer Optical Component Measuring Machine: a new Sub-nm Topography Measuring Device for X-ray Optics at BESSY
  • Siewert, Frank
  • SYNCHROTRON RADIATION INSTRUMENTATION: Eighth International Conference on Synchrotron Radiation Instrumentation, AIP Conference Proceedings https://doi.org/10.1063/1.1757928
conference January 2004
Optimal measurement strategies for effective suppression of drift errors journal November 2009
High precision tilt stage as a key element to a universal test mirror for characterization and calibration of slope measuring instruments journal May 2016
Aperture alignment in autocollimator-based deflectometric profilometers journal May 2016
Linear chirped slope profile for spatial calibration in slope measuring deflectometry journal May 2016
Multiplexed high resolution soft x-ray RIXS
  • Chuang, Y. -D.; Anderson, C.; Benk, M.
  • PROCEEDINGS OF THE 12TH INTERNATIONAL CONFERENCE ON SYNCHROTRON RADIATION INSTRUMENTATION – SRI2015, AIP Conference Proceedings https://doi.org/10.1063/1.4952931
conference January 2016
Grating metrology for X-ray and V-UV synchrotron beamlines at SOLEIL journal February 2019
Development of a long trace profiler in situ at National Synchrotron Radiation Research Center journal February 2019
Environmental influences on autocollimator-based angle and form metrology journal February 2019
Ex situ metrology and data analysis for optimization of beamline performance of aspherical pre-shaped x-ray mirrors at the advanced light source journal February 2019
Optimization of the size and shape of the scanning aperture in autocollimator-based deflectometric profilometers journal February 2019
Capabilities and limitations of the self-calibration of angle encoders journal March 2014
The new PTB angle comparator journal July 1998
On the characterization of ultra-precise X-ray optical components: advances and challenges in ex situ metrology journal August 2014
Developmental long-trace profiler using optimally aligned mirror-based pentaprism journal May 2011
Optimal alignment of mirror-based pentaprisms for scanning deflectometric devices journal July 2011
Advanced environmental control as a key component in the development of ultrahigh accuracy ex situ metrology for x-ray optics journal October 2015
XUV synchrotron optical components for the Advanced Light Source: summary of the requirements and the developmental program conference January 1993
Correlation analysis of surface slope metrology measurements of high quality x-ray optics conference September 2013
The developmental long trace profiler (DLTP) optimized for metrology of side-facing optics at the ALS conference September 2014
A new x-ray optics laboratory (XROL) at the ALS: mission, arrangement, metrology capabilities, performance, and future plans conference September 2014
New operational mode of the pencil beam interferometry based LTP conference September 2016
Development of a long trace profiler at SPring-8 using a newly developed slope sensor conference September 2016
Innovative nano-accuracy surface profiler for sub-50 nrad rms mirror test
  • Qian, Shinan; Idir, Mourad
  • Eighth International Symposium on Advanced Optical Manufacturing and Testing Technology (AOMATT2016), SPIE Proceedings https://doi.org/10.1117/12.2247575
conference October 2016
Development of a high performance surface slope measuring system for two-dimensional mapping of x-ray optics conference September 2017
New twist in the optical schematic of surface slope measuring long trace profiler conference September 2017
Correlation methods in optical metrology with state-of-the-art x-ray mirrors
  • Yashchuk, Valeriy V.; Centers, Gary; Gevorkyan, Gevork S.
  • Correlation Optics 2017, Thirteenth International Conference on Correlation Optics https://doi.org/10.1117/12.2305441
conference January 2018
The ALS OSMS: Optical Surface Measuring System for high accuracy two-dimensional slope metrology with state-of-the-art x-ray mirrors conference November 2018
Investigation on lateral resolution of surface slope profilers conference September 2019
Error reduction techniques for measuring long synchrotron mirrors conference November 1998
Mirror metrology and bender characterization at ESRF conference August 2005
Positioning errors of pencil-beam interferometers for long trace profilers conference August 2006
Air convection noise of pencil-beam interferometer for long trace profiler conference August 2006
Optimized use and calibration of autocollimators in deflectometry conference September 2007
Proposal for a universal test mirror for characterization of slope measuring instruments conference September 2007
Performance of the upgraded LTP-II at the ALS Optical Metrology Laboratory conference August 2008
Characterization of optical surfaces for the present generations of synchrotron sources conference October 2008
Requirements on hard x-ray grazing incidence optics for European XFEL: analysis and simulation of wavefront transformations conference May 2009
Automated suppression of errors in LTP-II slope measurements with x-ray optics conference September 2011
Pencil Beam Interferometer For Aspherical Optical Surfaces conference November 1982
Design Of A Long Trace Surface Profiler conference April 1987
Long Trace Profile Measurements On Cylindrical Aspheres conference January 1989
Interferometry of wave fronts reflected off conical surfaces journal January 1983
Improvements in the accuracy and the repeatability of long trace profiler measurements journal January 1999
Ultra-precise characterization of LCLS hard X-ray focusing mirrors by high resolution slope measuring deflectometry journal January 2012
Recent Developments in UV Optics for Ultra-Short, Ultra-Intense Coherent Light Sources journal January 2015

Similar Records

Optimization of the size and shape of the scanning aperture in autocollimator-based deflectometric profilometers
Journal Article · Mon Feb 04 19:00:00 EST 2019 · Review of Scientific Instruments · OSTI ID:1542362

Aperture alignment in autocollimator-based deflectometric profilometers
Journal Article · Sun May 15 00:00:00 EDT 2016 · Review of Scientific Instruments · OSTI ID:22597927

Aperture alignment in autocollimator-based deflectometric profilometers
Journal Article · Mon May 23 20:00:00 EDT 2016 · Review of Scientific Instruments · OSTI ID:1530235