Optimization of the size and shape of the scanning aperture in autocollimator-based deflectometric profilometers
Journal Article
·
· Review of Scientific Instruments
- Lawrence Berkeley National Lab. (LBNL), Berkeley, CA (United States)
- Physikalisch-Technische Bundesanstalt (Germany)
- Helmholtz Zentrum Berlin für Materialien und Energie, Berlin (Germany)
- Leibniz Inst. of Surface Engineering (IOM (Germany); Technische Universitaet Dresden (Germany)
- Leibniz Inst. of Surface Engineering (IOM (Germany)
Deflectometric profilometers based on industrial electronic autocollimators (ACs), as the ELCOMAT-3000, have become indispensable tools for precision form measurements of optical surfaces. A growing number of labs at synchrotron and free electron laser x-ray facilities are going for BESSY-II NOM-like versions of the AC-based profilometers. These tools have proven capable of characterizing state-of-the-art aspherical x-ray optics with an accuracy on the level of 100 nrad (root-mean-square) over the spatial frequency range limited by the size of the aperture used in the profilometer. Normally, a round aperture with a diameter of about 2.5 mm is used. Previous investigations have shown that with the optimally aligned 2.5-mm aperture, the spatial resolution of a NOM-like profilometer corresponding to the first zero-crossing of the optical transform function (OTF) is ~1.2 mm. In this paper, we investigate the performance of an AC ELCOMAT-3000 for a slope profilometer with different aperture sizes and shapes. The results of angular calibration of the AC equipped with circular and rectangular apertures placed at different distances from the AC are discussed. The calibration was performed at the Physikalisch-Technische Bundesanstalt using the original experimental arrangements, also discussed in the paper. The OTF measurements with the specially developed test sample with chirped surface slope profiles were performed at the Advanced Light Source X-Ray Optics Laboratory (XROL) in application to a new optical surface measuring system under development at the XROL. In the OTF measurements, we have shown that application of a rectangular aperture with dimensions of 1.5 mm × 3 mm improves the spatial resolution in the tangential direction by a factor of ~1.4 compared to that of the standard circular aperture of 2.5-mm diameter. We postulate that the conclusions of our investigations are crucial for reaching fundamental metrological limits in deflectometric profilometry utilizing state-of-the-art electronic autocollimators.
- Research Organization:
- Lawrence Berkeley National Laboratory (LBNL), Berkeley, CA (United States)
- Sponsoring Organization:
- USDOE Office of Science (SC)
- Grant/Contract Number:
- AC02-05CH11231
- OSTI ID:
- 1542362
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 2 Vol. 90; ISSN 0034-6748
- Publisher:
- American Institute of Physics (AIP)Copyright Statement
- Country of Publication:
- United States
- Language:
- English
The 6th International Workshop on X-ray Optics and Metrology—IWXM 2018
|
journal | February 2019 |
Transfer of autocollimator calibration for use with scanning gantry profilometers for accurate determination of surface slope and curvature of state-of-the-art x-ray mirrors
|
conference | October 2019 |
Similar Records
Aperture alignment in autocollimator-based deflectometric profilometers
Aperture alignment in autocollimator-based deflectometric profilometers
Investigation on lateral resolution of surface slope profilers
Journal Article
·
Sun May 15 00:00:00 EDT 2016
· Review of Scientific Instruments
·
OSTI ID:22597927
Aperture alignment in autocollimator-based deflectometric profilometers
Journal Article
·
Mon May 23 20:00:00 EDT 2016
· Review of Scientific Instruments
·
OSTI ID:1530235
Investigation on lateral resolution of surface slope profilers
Journal Article
·
Sun Sep 08 20:00:00 EDT 2019
· Proceedings of SPIE - The International Society for Optical Engineering
·
OSTI ID:1608216