Evaluation of Guardbanding Methods for Calibration and Product Acceptance
- Sandia National Lab. (SNL-NM), Albuquerque, NM (United States)
The Primary Standards Lab employs guardbanding methods to reduce risk of false acceptance in calibration when test uncertainty ratios are low. Similarly, production agencies guardband their requirements to reduce false accept rates in product acceptance. The root-sum-square guardbanding method is recommended by PSL, but many other guardbanding methods have been proposed in literature or implemented in commercial software. This report analyzes the false accept and reject rates resulting from the most common guardbanding methods. It is shown that the root-sum-square method and the Dobbert Managed Guardband strategy are similar and both are suitable for calibration and product acceptance work in the NSE.