Non-Invasive Functional-Brain-Imaging with an OPM-based Magnetoencephalography System

Journal Article · · PLoS ONE
 [1];  [2];  [2];  [2];  [3];  [4];  [5];  [6];  [2]
  1. Sandia National Lab. (SNL-NM), Albuquerque, NM (United States); DOE/OSTI
  2. Sandia National Lab. (SNL-NM), Albuquerque, NM (United States)
  3. Candoo Systems Inc., Coquitlam, BC (Canada)
  4. StimScience, Inc., Berkeley, CA (United States)
  5. Univ. of Washington, Seattle, WA (United States)
  6. The Mind Research Network and Lovelace Biomedical and Environmental Research Inst., Albuquerque, NM (United States)

A non-invasive functional-brain-imaging system based on optically-pumped-magnetometers (OPM) is presented. The OPM-based magnetoencephalography (MEG) system features 20 OPM channels conforming to the subject’s scalp. We have conducted two MEG experiments on three subjects: assessment of somatosensory evoked magnetic field (SEF) and auditory evoked magnetic field (AEF) using our OPM-based MEG system and a commercial MEG system based on superconducting quantum interference devices (SQUIDs). We cross validated the robustness of our system by calculating the distance between the location of the equivalent current dipole (ECD) yielded by our OPM-based MEG system and the ECD location calculated by the commercial SQUID-based MEG system. We achieved sub-centimeter accuracy for both SEF and AEF responses in all three subjects. Due to the proximity (12 mm) of the OPM channels to the scalp, it is anticipated that future OPM-based MEG systems will offer enhanced spatial resolution as they will capture finer spatial features compared to traditional MEG systems employing SQUIDs.

Research Organization:
Sandia National Laboratories (SNL-NM), Albuquerque, NM (United States)
Sponsoring Organization:
USDOE Office of Science (SC), Biological and Environmental Research (BER). Biological Systems Science Division
Grant/Contract Number:
NA0003525
OSTI ID:
1628983
Journal Information:
PLoS ONE, Journal Name: PLoS ONE Journal Issue: 1 Vol. 15; ISSN 1932-6203
Publisher:
Public Library of ScienceCopyright Statement
Country of Publication:
United States
Language:
English

References (39)

Validation of a method for coregistering scalp recording locations with 3D structural MR images journal November 2008
Requirements for Coregistration Accuracy in On-Scalp MEG journal June 2018
Measuring MEG closer to the brain: Performance of on-scalp sensor arrays journal February 2017
A new generation of magnetoencephalography: Room temperature measurements using optically-pumped magnetometers journal April 2017
The impact of improved MEG–MRI co-registration on MEG connectivity analysis journal August 2019
A subfemtotesla multichannel atomic magnetometer journal April 2003
Moving magnetoencephalography towards real-world applications with a wearable system journal March 2018
Unshielded three-axis vector operation of a spin-exchange-relaxation-free atomic magnetometer journal November 2004
Presentation of electromagnetic multichannel data: The signal space separation method journal June 2005
BabyMEG: A whole-head pediatric magnetoencephalography system for human brain development research journal September 2016
The magnetic lead field theorem in the quasi-static approximation and its use for magnetoencephalography forward calculation in realistic volume conductors journal October 2003
Magnetoencephalography—theory, instrumentation, and applications to noninvasive studies of the working human brain journal April 1993
Artifact Removal in Magnetoencephalogram Background Activity With Independent Component Analysis journal November 2007
Magnetoencephalography: Detection of the Brain's Electrical Activity with a Superconducting Magnetometer journal February 1972
Four-channel optically pumped atomic magnetometer for magnetoencephalography journal January 2016
Magnetoencephalography of Epilepsy with a Microfabricated Atomic Magnetrode journal October 2014
A Compact, High Performance Atomic Magnetometer for Biomedical Applications text January 2013
A 20-Channel Magnetoencephalography System Based on Optically Pumped Magnetometers text January 2017
Signal-space projection method for separating MEG or EEG into components journal March 1997
Detection of very weak magnetic fields (10−9gauss) by 87Rb zero-field level crossing resonances journal February 1969
Optical Pumping journal October 1960
Magnetoencephalography with an atomic magnetometer journal November 2006
Cross-validation of microfabricated atomic magnetometers with superconducting quantum interference devices for biomagnetic applications journal September 2010
Magnetoencephalography with a two-color pump-probe, fiber-coupled atomic magnetometer journal December 2010
Spatiotemporal signal space separation method for rejecting nearby interference in MEG measurements journal March 2006
A compact, high performance atomic magnetometer for biomedical applications journal November 2013
Fetal magnetocardiography measurements with an array of microfabricated optically pumped magnetometers journal June 2015
A 20-channel magnetoencephalography system based on optically pumped magnetometers journal November 2017
Modified activation of somatosensory cortical network in patients with right-hemisphere stroke journal October 1999
Experimental Study of Zeeman Light Shifts in Weak Magnetic Fields journal February 1972
Optical Pumping journal April 1972
Magnetoencephalography—theory, instrumentation, and applications to noninvasive studies of the working human brain journal April 1993
Artifact Removal in Magnetoencephalogram Background Activity With Independent Component Analysis journal November 2007
Optical Pumping journal October 1960
Optical Pumping journal August 1963
FieldTrip: Open Source Software for Advanced Analysis of MEG, EEG, and Invasive Electrophysiological Data journal January 2011
Four-channel optically pumped atomic magnetometer for magnetoencephalography journal January 2016
Combined EEG/MEG Can Outperform Single Modality EEG or MEG Source Reconstruction in Presurgical Epilepsy Diagnosis journal March 2015
Magnetoencephalography of Epilepsy with a Microfabricated Atomic Magnetrode journal October 2014

Similar Records

A 20-channel magnetoencephalography system based on optically pumped magnetometers
Journal Article · Mon Oct 16 00:00:00 EDT 2017 · Physics in Medicine and Biology · OSTI ID:1429765

Single-trial classification of evoked responses to auditory tones using OPM- and SQUID-MEG
Journal Article · Fri Oct 06 00:00:00 EDT 2023 · Journal of Neural Engineering · OSTI ID:2007979

Related Subjects