Implementation of a Talbot–Lau x-ray deflectometer diagnostic platform for the OMEGA EP laser
Journal Article
·
· Review of Scientific Instruments
- Johns Hopkins Univ., Baltimore, MD (United States); Johns Hopkins University
- Johns Hopkins Univ., Baltimore, MD (United States)
- Univ. of Rochester, NY (United States). Lab. for Laser Energetics
- General Atomics, San Diego, CA (United States)
- Microworks GmbH, Karlsruhe (Germany)
- Los Alamos National Lab. (LANL), Los Alamos, NM (United States)
- Sandia National Lab. (SNL-NM), Albuquerque, NM (United States)
- Univ. of Michigan, Ann Arbor, MI (United States)
A Talbot–Lau X-ray Deflectometer (TXD) was implemented in the OMEGA EP laser facility to characterize the evolution of an irradiated foil ablation front by mapping electron densities >1022 cm-3 by means of Moiré deflectometry. The experiment used a short-pulse laser (30–100 J, 10 ps) and a foil copper target as an x-ray backlighter source. In the first experimental tests performed to benchmark the diagnostic platform, grating survival was demonstrated and x-ray backlighter laser parameters that deliver Moiré images were described. Finally, the necessary modifications to accurately probe the ablation front through TXD using the EP-TXD diagnostic platform are discussed.
- Research Organization:
- Johns Hopkins Univ., Baltimore, MD (United States); Los Alamos National Laboratory (LANL), Los Alamos, NM (United States)
- Sponsoring Organization:
- USDOE National Nuclear Security Administration (NNSA)
- Grant/Contract Number:
- NA0003882; NA0002955; NA0003526; NA0002956; 89233218CNA000001
- OSTI ID:
- 1604067
- Alternate ID(s):
- OSTI ID: 1650622
OSTI ID: 1781576
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 2 Vol. 91; ISSN 0034-6748
- Publisher:
- American Institute of Physics (AIP)Copyright Statement
- Country of Publication:
- United States
- Language:
- English
Proof-of-concept Talbot–Lau x-ray interferometry with a high-intensity, high-repetition-rate, laser-driven K-alpha source
|
journal | January 2020 |
Similar Records
Talbot-Lau x-ray deflectometer: Refraction-based HEDP imaging diagnostic
Journal Article
·
Fri Jun 18 00:00:00 EDT 2021
· Review of Scientific Instruments
·
OSTI ID:1837686