Measured accuracy and precision in quantitative SPECT imaging with iodine-123

Journal Article · · IEEE Transactions on Nuclear Science
DOI:https://doi.org/10.1109/23.467738· OSTI ID:148058
; ; ; ;  [1]
  1. Duke Univ. Medical Center, Durham, NC (United States). Dept. of Radiology
This study evaluated {sup 123}I SPECT volume and activity quantitation in terms of accuracy and precision using experimental phantom data. Two reconstruction methods were considered. The filtered back projection (FB) method with 2D Metz pre-filtering and single iteration Chang attenuation compensation was compared with an iterative ML-EM method that used Gaussian post-filtering and incorporated the 3D divergent detector response and attenuation in the detection kernel. Both methods were examined over a range of resolution and noise characteristics. An ensemble of 21 SPECT data sets were acquired of a cylindrical phantom that contained spheres of size 21.4, 12.0, and 5.7 ml, filled with high purity {sup 123}I in a uniform background (5:1 uptake ratio). Each data set was reconstructed by both methods, and sphere volume and activity measurements were obtained from the reconstructions using a semi-automatic method that defines the region-of-interest based on the intensity gradient in the area of increased {sup 123}I concentration. The mean and standard deviation of the 21 measurements provided accuracy and precision indices, respectively. For the 21.4 ml sphere, the measured volume with either reconstruction method was accurate to within 5% of the true volume with a standard deviation of approximately 8% of the true volume. For the 12.0 ml sphere the measured volume with the iterative method exhibited substantially smaller standard deviation for equal accuracy compared with the FB method. Sphere activity was underestimated with both methods due to spatial resolution effects. The 5.7 ml sphere was generally below the size limit for volume and activity quantitation by these method.
OSTI ID:
148058
Report Number(s):
CONF-941061--
Journal Information:
IEEE Transactions on Nuclear Science, Journal Name: IEEE Transactions on Nuclear Science Journal Issue: 4Pt1 Vol. 42; ISSN 0018-9499; ISSN IETNAE
Country of Publication:
United States
Language:
English

Similar Records

Quantitative SPECT brain imaging: Effects of attenuation and detector response
Journal Article · Tue Jun 01 00:00:00 EDT 1993 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:6309063

Quantitative SPECT reconstruction of iodine-123 data
Journal Article · Thu Feb 28 23:00:00 EST 1991 · Journal of Nuclear Medicine; (USA) · OSTI ID:6042239

Volume and activity quantitation with iodine-123 SPECT
Journal Article · Sat Oct 01 00:00:00 EDT 1994 · Journal of Nuclear Medicine · OSTI ID:458867