Unabsorbed light beamlets for diagnosing cross-beam energy transfer
Journal Article
·
· Review of Scientific Instruments
- Univ. of Rochester, Rochester, NY (United States). Lab. for Laser Energetics; Laboratory for Laser Energetics, University of Rochester
- Univ. of Rochester, Rochester, NY (United States). Lab. for Laser Energetics
Here, a new diagnostic has been fielded on OMEGA to diagnose cross-beam energy transfer (CBET) during direct-drive implosions. Unabsorbed light from each OMEGA laser beam is imaged as a distinct “spot” onto a gated optical imager. Each spot is in essence the end point of a beamlet of light that originates from different regions of each beam profile and follows a path determined by refraction. The intensity of light in the beamlet varies along its path as a result of absorption and CBET with other beamlets. This diagnostic allows the investigation of the effects of CBET on laser energy from specific locations of the beam profile. The diagnostic records images in two 200-ps time windows and includes a Wollaston prism to split each beamlet into two orthogonal polarizations recorded on separate images, allowing the absolute polarization of each beamlet to be determined. This diagnostic has provided the first evidence of polarization rotation caused by CBET during direct-drive implosions.
- Research Organization:
- Univ. of Rochester, Rochester, NY (United States). Lab. for Laser Energetics
- Sponsoring Organization:
- USDOE; USDOE National Nuclear Security Administration (NNSA)
- Grant/Contract Number:
- NA0001944
- OSTI ID:
- 1460094
- Alternate ID(s):
- OSTI ID: 1459476
- Report Number(s):
- 2017-258, 2374; 2017-258, 2374, 1416
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 10 Vol. 89; ISSN 0034-6748
- Publisher:
- American Institute of Physics (AIP)Copyright Statement
- Country of Publication:
- United States
- Language:
- English
Similar Records
Unabsorbed light beamlets for diagnosing coronal density profiles and absorption nonuniformity in direct-drive implosions on OMEGA
Polarimetry diagnostic on OMEGA EP using a 10-ps, 263-nm probe beam
Journal Article
·
Tue Apr 06 20:00:00 EDT 2021
· Review of Scientific Instruments
·
OSTI ID:1774750
Polarimetry diagnostic on OMEGA EP using a 10-ps, 263-nm probe beam
Journal Article
·
Mon Aug 11 00:00:00 EDT 2014
· Review of Scientific Instruments
·
OSTI ID:1156677