Responsivity calibration of the LoWEUS spectrometer

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4960729· OSTI ID:1349666
 [1];  [2];  [3];  [3];  [3]
  1. Univ. of California, Berkeley, CA (United States)
  2. Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
  3. Princeton Plasma Physics Lab. (PPPL), Princeton, NJ (United States)
We performed an in situ calibration of the relative responsivity function of the Long-Wavelength Extreme Ultraviolet Spectrometer (LoWEUS), while operating on the Lithium Tokamak Experiment (LTX) at Princeton Plasma Physics Laboratory. The calibration was accomplished by measuring oxygen lines, which are typically present in LTX plasmas. The measured spectral line intensities of each oxygen charge state were then compared to the calculated emission strengths given in the CHIANTI atomic database. Normalizing the strongest line in each charge state to the CHIANTI predictions, we obtained the differences between the measured and predicted values for the relative strengths of the other lines of a given charge state. We find that a 3rd degree polynomial function provides a good fit to the data points. Lastly, our measurements show that the responsivity between about 120 and 300 Å varies by factor of ~30.
Research Organization:
Lawrence Livermore National Laboratory (LLNL), Livermore, CA (United States); Princeton Plasma Physics Laboratory (PPPL), Princeton, NJ (United States)
Sponsoring Organization:
USDOE
Grant/Contract Number:
AC02-09CH11466; AC52-07NA27344
OSTI ID:
1349666
Alternate ID(s):
OSTI ID: 1400102
OSTI ID: 22596393
Report Number(s):
LLNL-JRNL--698642
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 11 Vol. 87; ISSN 0034-6748; ISSN RSINAK
Publisher:
American Institute of Physics (AIP)Copyright Statement
Country of Publication:
United States
Language:
English

References (10)

CHIANTI - an atomic database for emission lines: I. Wavelengths greater than 50 Å journal October 1997
Photometric calibration of an EUV flat field spectrometer at the advanced light source journal March 2003
Vacuum ultraviolet impurity spectroscopy on the Alcator C-Mod tokamak journal October 2010
High-resolution time-resolved extreme ultraviolet spectroscopy on NSTX journal October 2012
Emission‐Line Spectra of Ar ix –Ar xvi in the Soft X‐Ray Region 20–50 A journal June 2003
Emission‐Line Spectra of S vii –S xiv in the 20–75 A Wavelength Region journal June 2005
Chianti—An Atomic Database for Emission Lines. Xiii. soft X-Ray Improvements and Other Changes journal January 2013
EUV spectroscopy on NSTX journal July 2010
Laboratory astrophysics in the extreme ultraviolet journal January 2008
Measuring plasma impurities in Alcator C-Mod as a function of time in the extreme ultraviolet 1 This article is part of a Special Issue on the 10th International Colloquium on Atomic Spectra and Oscillator Strengths for Astrophysical and Laboratory Plasmas. journal May 2011

Cited By (2)

Electron-density-sensitive Line Ratios of Fe xiii– xvi from Laboratory Sources Compared to CHIANTI journal February 2018
EUV Spectra and Line Ratios of Multiply Ionized Calcium and Argon Atoms in a Laboratory Plasma journal October 2018

Similar Records

Responsivity calibration of the LoWEUS spectrometer
Journal Article · Mon Nov 14 23:00:00 EST 2016 · Review of Scientific Instruments · OSTI ID:22596393

Absolute calibration of a time-resolved high resolution x-ray spectrometer for the National Ignition Facility (invited)
Journal Article · Thu Oct 18 20:00:00 EDT 2018 · Review of Scientific Instruments · OSTI ID:1512496

Electron-density-sensitive Line Ratios of Fe xiii– xvi from Laboratory Sources Compared to CHIANTI
Journal Article · Wed Feb 14 19:00:00 EST 2018 · The Astrophysical Journal (Online) · OSTI ID:1432662