We performed an in situ calibration of the relative responsivity function of the Long-Wavelength Extreme Ultraviolet Spectrometer (LoWEUS), while operating on the Lithium Tokamak Experiment (LTX) at Princeton Plasma Physics Laboratory. The calibration was accomplished by measuring oxygen lines, which are typically present in LTX plasmas. The measured spectral line intensities of each oxygen charge state were then compared to the calculated emission strengths given in the CHIANTI atomic database. Normalizing the strongest line in each charge state to the CHIANTI predictions, we obtained the differences between the measured and predicted values for the relative strengths of the other lines of a given charge state. We find that a 3rd degree polynomial function provides a good fit to the data points. Lastly, our measurements show that the responsivity between about 120 and 300 Å varies by factor of ~30.
- Research Organization:
- Lawrence Livermore National Laboratory (LLNL), Livermore, CA (United States); Princeton Plasma Physics Laboratory (PPPL), Princeton, NJ (United States)
- Sponsoring Organization:
- USDOE
- Grant/Contract Number:
- AC02-09CH11466; AC52-07NA27344
- OSTI ID:
- 1349666
- Alternate ID(s):
- OSTI ID: 1400102
OSTI ID: 22596393
- Report Number(s):
- LLNL-JRNL--698642
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 11 Vol. 87; ISSN 0034-6748; ISSN RSINAK
- Publisher:
- American Institute of Physics (AIP)Copyright Statement
- Country of Publication:
- United States
- Language:
- English
Electron-density-sensitive Line Ratios of Fe xiii– xvi from Laboratory Sources Compared to CHIANTI
|
journal | February 2018 |
EUV Spectra and Line Ratios of Multiply Ionized Calcium and Argon Atoms in a Laboratory Plasma
|
journal | October 2018 |
Similar Records
Responsivity calibration of the LoWEUS spectrometer
Absolute calibration of a time-resolved high resolution x-ray spectrometer for the National Ignition Facility (invited)
Electron-density-sensitive Line Ratios of Fe xiii– xvi from Laboratory Sources Compared to CHIANTI
Journal Article
·
Mon Nov 14 23:00:00 EST 2016
· Review of Scientific Instruments
·
OSTI ID:22596393
Absolute calibration of a time-resolved high resolution x-ray spectrometer for the National Ignition Facility (invited)
Journal Article
·
Thu Oct 18 20:00:00 EDT 2018
· Review of Scientific Instruments
·
OSTI ID:1512496
Electron-density-sensitive Line Ratios of Fe xiii– xvi from Laboratory Sources Compared to CHIANTI
Journal Article
·
Wed Feb 14 19:00:00 EST 2018
· The Astrophysical Journal (Online)
·
OSTI ID:1432662