Development and implementation of a portable grating interferometer system as a standard tool for testing optics at the Advanced Photon Source beamline 1-BM
Journal Article
·
· Review of Scientific Instruments
- Argonne National Lab. (ANL), Lemont, IL (United States). Advanced Photon Source (APS)
We developed a portable X-ray grating interferometer setup as a standard tool for testing optics at the Advanced Photon Source (APS) beamline 1-BM. The interferometer can be operated in phase-stepping, Moiré, or single-grating harmonic imaging mode with 1-D or 2-D gratings. All of the interferometer motions are motorized; hence, it is much easier and quicker to switch between the different modes of operation. A novel aspect of this new instrument is its designed portability. While the setup is designed to be primarily used as a standard tool for testing optics at 1-BM, it could be potentially deployed at other APS beamlines for beam coherence and wavefront characterization or imaging. Herein, the design of the interferometer system is described in detail and coherence measurements obtained at the APS 34-ID-E beamline are presented. The coherence was probed in two directions using a 2-D checkerboard, a linear, and a circular grating at X-ray energies of 8 keV, 11 keV, and 18 keV.
- Research Organization:
- Argonne National Lab. (ANL), Argonne, IL (United States). Advanced Photon Source (APS)
- Sponsoring Organization:
- USDOE Office of Science (SC), Basic Energy Sciences (BES) (SC-22). Scientific User Facilities Division
- Grant/Contract Number:
- AC02-06CH11357
- OSTI ID:
- 1332986
- Alternate ID(s):
- OSTI ID: 22597932
OSTI ID: 1254051
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 5 Vol. 87; ISSN 0034-6748; ISSN RSINAK
- Publisher:
- American Institute of Physics (AIP)Copyright Statement
- Country of Publication:
- United States
- Language:
- English
Similar Records
Development and implementation of a portable grating interferometer system as a standard tool for testing optics at the Advanced Photon Source beamline 1-BM
X-ray Optics Testing Beamline 1-BM at the Advanced Photon Source
X-ray optics testing beamline 1-BM at the advanced photon source
Journal Article
·
Sun May 15 00:00:00 EDT 2016
· Review of Scientific Instruments
·
OSTI ID:22597932
X-ray Optics Testing Beamline 1-BM at the Advanced Photon Source
Conference
·
Wed Jul 27 00:00:00 EDT 2016
·
OSTI ID:1336950
X-ray optics testing beamline 1-BM at the advanced photon source
Journal Article
·
Wed Jul 27 00:00:00 EDT 2016
· AIP Conference Proceedings
·
OSTI ID:22608366