Composite photon-electron spectrometer from NaI(T1) crystals and plastic scintillators
Journal Article
·
· Instruments and Experimental Techniques
OSTI ID:133127
- P.N. Lebedev Institute of Physics, Moscow (Russian Federation); and others
The structure of a composite photon-electron spectrometer based on NaI(T1) crystals and plastic scintillators is described. A spectrometer calibration under a beam of secondary electrons with an energy of 50-150 MeV is described. At a work-load of {approximately}10{sup 3} Hz energy resolution is {sigma}/E = 0.03E{sup -0.25}, and the time resolution is {approximately}8 nsec.
- OSTI ID:
- 133127
- Journal Information:
- Instruments and Experimental Techniques, Journal Name: Instruments and Experimental Techniques Journal Issue: 1 Vol. 38; ISSN 0020-4412; ISSN INETAK
- Country of Publication:
- United States
- Language:
- English
Similar Records
THE EFFECT OF COLLIMATION ON THE RESOLUTION AND PEAK-TO-TOTAL RATIO OF A NaI(T1) CRYSTAL
GAMMA SPECTROMETER WITH ORGANIC SCINTILLATOR CAPABLE OF TIME RESOLVING SELECTION OF $gamma$ RADIATION
NaI (T1) scintillation with counter with vacuum photodiode
Technical Report
·
Tue Jan 30 23:00:00 EST 1962
·
OSTI ID:4786630
GAMMA SPECTROMETER WITH ORGANIC SCINTILLATOR CAPABLE OF TIME RESOLVING SELECTION OF $gamma$ RADIATION
Journal Article
·
Tue Feb 28 23:00:00 EST 1961
· Pribory i Tekh. Ekspt.
·
OSTI ID:4840309
NaI (T1) scintillation with counter with vacuum photodiode
Journal Article
·
Mon Sep 01 00:00:00 EDT 1986
· Instrum. Exp. Tech. (Engl. Transl.); (United States)
·
OSTI ID:6863214