Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Composite photon-electron spectrometer from NaI(T1) crystals and plastic scintillators

Journal Article · · Instruments and Experimental Techniques
OSTI ID:133127
; ;  [1]
  1. P.N. Lebedev Institute of Physics, Moscow (Russian Federation); and others
The structure of a composite photon-electron spectrometer based on NaI(T1) crystals and plastic scintillators is described. A spectrometer calibration under a beam of secondary electrons with an energy of 50-150 MeV is described. At a work-load of {approximately}10{sup 3} Hz energy resolution is {sigma}/E = 0.03E{sup -0.25}, and the time resolution is {approximately}8 nsec.