Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Validation of GC and HPLC systems for residue studies

Conference ·
OSTI ID:126059
 [1]
  1. Horizon Labs., Inc., Columbia, MO (United States)

For residue studies, GC and HPLC system performance must be validated prior to and during use. One excellent measure of system performance is the standard curve and associated chromatograms used to construct that curve. The standard curve is a model of system response to an analyte over a specific time period, and is prima facia evidence of system performance beginning at the auto sampler and proceeding through the injector, column, detector, electronics, data-capture device, and printer/plotter. This tool measures the performance of the entire chromatographic system; its power negates most of the benefits associated with costly and time-consuming validation of individual system components. Other measures of instrument and method validation will be discussed, including quality control charts and experimental designs for method validation.

OSTI ID:
126059
Report Number(s):
CONF-950402--
Country of Publication:
United States
Language:
English

Similar Records

Development of an HPLC, GC/MS method for analysis of HYGAS oil samples
Technical Report · Fri Jun 01 00:00:00 EDT 1979 · OSTI ID:5769895

Validating data from automated PAMS GC systems
Conference · Fri Dec 30 23:00:00 EST 1994 · OSTI ID:197335

Internal Standards for Quantitative Analysis of Chemical Warfare Agents by the GC/MS Method: Nerve Agents