Evaluation of Two Guralp Preamplifiers for GS21 Seismometer Application

Technical Report ·
DOI:https://doi.org/10.2172/1213489· OSTI ID:1213489
 [1];  [1]
  1. Sandia National Lab. (SNL-NM), Albuquerque, NM (United States)

Sandia National Laboratories has tested and evaluated two Guralp preamplifiers for use with a GS21 seismometer application. The two preamplifiers have a gain factor of 61.39. The purpose of the preamplifier evaluation was to determine a measured gain factor, transfer function, total harmonic distortion, self-noise, application passband, dynamic range, seismometer calibration pass-through, and to comment on any issues encountered during the evaluation. The test results included in this report were in response to static, tonal, and dynamic input signals. The Guralp GS21 preamplifiers are being evaluated for potential use in the International Monitoring System (IMS) of the Comprehensive Nuclear Test-Ban-Treaty Organization (CTBTO). Test methodologies used were based on IEEE Standards 1057 for Digitizing Waveform Recorders and 1241 for Analog to Digital Converters

Research Organization:
Sandia National Laboratories (SNL-NM), Albuquerque, NM (United States)
Sponsoring Organization:
USDOE National Nuclear Security Administration (NNSA), Office of Defense Nuclear Nonproliferation
DOE Contract Number:
AC04-94AL85000
OSTI ID:
1213489
Report Number(s):
SAND--2015-7076; 603212
Country of Publication:
United States
Language:
English

Similar Records

Evaluation of Two Guralp Preamplifiers for GS13 Seismometer Application.
Technical Report · Fri Sep 01 00:00:00 EDT 2017 · OSTI ID:1398237

2017 Guralp Affinity Digitizer Evaluation.
Technical Report · Wed Feb 28 23:00:00 EST 2018 · OSTI ID:1425328

Next Generation Qualification: Guralp CMG-3V Seismometer Evaluation
Technical Report · Wed Aug 01 00:00:00 EDT 2018 · OSTI ID:1760392