Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Detection of ``single-leg separated`` heart valves using statistical pattern recognition with the nearest neighbor classifier

Technical Report ·
DOI:https://doi.org/10.2172/10177333· OSTI ID:10177333
The goal of this work was to detect ``single-leg separated`` Bjoerk-Shiley Convexo-Concave heart valves which had been implanted in sheep. A ``single-leg separated`` heart valve contains a fracture in the outlet strut resulting in an increased risk of mechanical failure. The approach presented in this report detects such fractures by applying statistical pattern recognition with the nearest neighbor classifier to the acoustic signatures of the valve opening. This approach is discussed and results of applying it to real data are given.
Research Organization:
Lawrence Livermore National Lab., CA (United States)
Sponsoring Organization:
USDOE, Washington, DC (United States)
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
10177333
Report Number(s):
UCRL-ID--114802; ON: DE93018911
Country of Publication:
United States
Language:
English