The design principles, construction and characterization of a 4- bounce dispersive crystal monochromator is discussed. This monochromator is designed to reduce the bandpass of synchrotron radiation to 10--50 meV level, without sacrificing angular acceptance. This is achieved by combining an asymmetrically-cut, low order reflection with a symmetrically-cut, high order reflection in a nested configuration. This monochromator is being used as a beam conditioner for nuclear resonant scattering of synchrotron radiation to produce x-rays with {mu}eV{minus}neV resolution in the hard x-ray regime.
- Research Organization:
- Argonne National Lab., IL (United States)
- Sponsoring Organization:
- USDOE, Washington, DC (United States)
- DOE Contract Number:
- W-31109-ENG-38
- OSTI ID:
- 10105728
- Report Number(s):
- ANL/XFD/CP-76595; CONF-920768-7; ON: DE93004195
- Resource Relation:
- Conference: Conference on physics and engineering of advanced detectors and collection/focusing systems,San Diego, CA (United States),19-24 Jul 1992; Other Information: PBD: Jun 1992
- Country of Publication:
- United States
- Language:
- English
Similar Records
High-energy resolution, high-angular acceptance crystal monochromator
High energy resolution, high angular acceptance crystal monochromator
High energy resolution, high angular acceptance crystal monochromator
Conference
·
Mon Jun 01 00:00:00 EDT 1992
·
OSTI ID:10105728
+1 more
High energy resolution, high angular acceptance crystal monochromator
Patent
·
Tue Jun 04 00:00:00 EDT 1996
·
OSTI ID:10105728
High energy resolution, high angular acceptance crystal monochromator
Patent
·
Tue Jun 04 00:00:00 EDT 1996
·
OSTI ID:10105728
Related Subjects
43 PARTICLE ACCELERATORS
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
MONOCHROMATORS
DESIGN
ADVANCED PHOTON SOURCE
SYNCHROTRON RADIATION
SCATTERING
X RADIATION
PERFORMANCE TESTING
IRON 57
MOESSBAUER EFFECT
430303
440101
EXPERIMENTAL FACILITIES AND EQUIPMENT
GENERAL DETECTORS OR MONITORS AND RADIOMETRIC INSTRUMENTS
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
MONOCHROMATORS
DESIGN
ADVANCED PHOTON SOURCE
SYNCHROTRON RADIATION
SCATTERING
X RADIATION
PERFORMANCE TESTING
IRON 57
MOESSBAUER EFFECT
430303
440101
EXPERIMENTAL FACILITIES AND EQUIPMENT
GENERAL DETECTORS OR MONITORS AND RADIOMETRIC INSTRUMENTS