Fast ion chambers for SLC
Conference
·
OSTI ID:10105385
Beam diagnostic ion chambers are used throughout the SLC to perform a variety of tasks including locating beam losses along the beam direction, determining localized losses from individual bunches in a multibunch beam, and detecting scattered particles from beam profile wire scanners where backgrounds are too high to use photomultiplier tubes. Construction and instrumentation of very fast ion chambers with pulse duration of less than 60ns are detailed. Long ion chambers referred to as PLIC (Panofsky`s Long Ion Chamber) are the primary diagnostic used to locate losses in all the SLC transport lines. Accurately locating beam loss with the use of fiducial cables and coaxial switches will be discussed.
- Research Organization:
- Stanford Linear Accelerator Center, Menlo Park, CA (United States)
- Sponsoring Organization:
- USDOE, Washington, DC (United States)
- DOE Contract Number:
- AC03-76SF00515
- OSTI ID:
- 10105385
- Report Number(s):
- SLAC-PUB--6296; CONF-930511--499; ON: DE94003439
- Country of Publication:
- United States
- Language:
- English
Similar Records
Long ion chamber systems for the SLC (Stanford Linear Collider)
PLIC signal voltages in the SLAC accelerator scaled to the SLC arcs
Wire scanners for beam size and emittance measurements at the SLC
Conference
·
Tue Feb 28 23:00:00 EST 1989
·
OSTI ID:6153453
PLIC signal voltages in the SLAC accelerator scaled to the SLC arcs
Technical Report
·
Mon Jun 16 00:00:00 EDT 1986
·
OSTI ID:5746668
Wire scanners for beam size and emittance measurements at the SLC
Conference
·
Wed May 01 00:00:00 EDT 1991
·
OSTI ID:5528761