skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: AUTOMATED ELEMENTAL ANALYSIS USING ENERGY DISPERSIVE X-RAY FLUORESCENCE ANALYSIS

Conference ·
OSTI ID:1004192

The application of energy dispersive x-ray fluorescence analysis to the measurements of air particulate samples has been described in several recent publications. Among its many advantages, the possibility of automatic analysis for large scale monitoring programs has not been widely discussed. We describe the completion of a program in which a total of 34,000 air particulate samples have been collected and analyzed over the past two years. An automatic photon-excited energy dispersive spectrometer was used to obtain elemental concentrations for 27 elements on each sample. The sensitivity of the system is adequate to detect quantities of 10 ngm/cm{sup 2} or less in a measurement time of approximately six minutes per sample. The precision and accuracy of the results will be compared to statistical analyses of the data set. Summaries of the results are presented and the implications for air particulate monitoring discussed.

Research Organization:
Lawrence Berkeley National Lab. (LBNL), Berkeley, CA (United States)
Sponsoring Organization:
Nuclear Science Division
DOE Contract Number:
DE-AC02-05CH11231
OSTI ID:
1004192
Report Number(s):
LBL-7211; TRN: US201103%%454
Resource Relation:
Conference: 4th Joint Conference on Sensing of Environmental Pollutants, New Orleans, LA, November 6-11, 1977
Country of Publication:
United States
Language: