Sub-millikelvin stabilization of a closed cycle cryocooler
Journal Article
·
· Review of Scientific Instruments
- Brookhaven National Lab. (BNL), Upton, NY (United States)
In this study, intrinsic temperature oscillations (with the amplitude up to 1 K) of a closed cycle cryocooler are stabilized by a simple thermal damping system. It employs three different materials with different thermal conductivity and capacity at various temperatures. The amplitude of oscillations of the sample temperature is reduced to less than 1 mK, in the temperature range from 4 K to 300 K, while the cooling power is virtually undiminished. The damping system is small, inexpensive, can be retrofitted to most existing closed cycle cryocoolers, and may improve measurements of any temperature-sensitive physics properties.
- Research Organization:
- Brookhaven National Laboratory (BNL), Upton, NY (United States)
- Sponsoring Organization:
- USDOE Office of Science (SC), Basic Energy Sciences (BES)
- Grant/Contract Number:
- SC00112704
- OSTI ID:
- 1182495
- Report Number(s):
- BNL-107390-2015-JA; RSINAK; R&D Project: MA509MACA; KC0203020
- Journal Information:
- Review of Scientific Instruments, Vol. 85, Issue 10; ISSN 0034-6748
- Publisher:
- American Institute of Physics (AIP)Copyright Statement
- Country of Publication:
- United States
- Language:
- English
Cited by: 20 works
Citation information provided by
Web of Science
Web of Science
Dependence of the critical temperature in overdoped copper oxides on superfluid density
|
journal | August 2016 |
Realization of an ultra-high precision temperature control in a cryogen-free cryostat
|
journal | October 2018 |
Influence of transverse stress exerted at room temperature on the superconducting properties of Nb 3 Sn wires
|
journal | August 2019 |
Search for ferromagnetic order in overdoped copper-oxide superconductors
|
journal | April 2017 |
Similar Records
10 deg K triple-expansion Stirling-cycle cryocooler
Closed cycle cryocooler for low temperature electronic circuits. Phase II. Preliminary design. Final report 15 May-30 Sep 81
Closed cycle cryocooler with hybrid heat exchanger-regenerator: Phase 1 final report
Technical Report
·
Thu Dec 01 00:00:00 EST 1983
·
OSTI ID:1182495
Closed cycle cryocooler for low temperature electronic circuits. Phase II. Preliminary design. Final report 15 May-30 Sep 81
Technical Report
·
Sun Aug 01 00:00:00 EDT 1982
·
OSTI ID:1182495
Closed cycle cryocooler with hybrid heat exchanger-regenerator: Phase 1 final report
Technical Report
·
Fri Feb 24 00:00:00 EST 1984
·
OSTI ID:1182495