The purpose of this report is to investigate the possibility of an embryo to receive a dose of more than 10 cGy, the threshold of malformation induction in embryos reported by the International Commission on Radiological Protection, during barium enema examinations. Thermoluminescent dosimeters were place in a phantom to calculate the depth-to-skin conversion coefficient needed for dose estimation at the average embryo depth in patients. Barium enema examinations were performed in 20 women of childbearing age with diagnostic problems demanding longer fluoroscopy times. Doses at 6 cm, the average embryo depth, were determined by measurements at the patients` skin followed by dose calculation at the site of interest. The range of doses estimated at embryo depth for patients was 1.9 to 8.2 cGy. The dose always exceeded 5 cGy when fluoroscopy time was longer than 7 minutes. The dose at the embryo depth never exceeded 10 cGy. This study indicates that fluoroscopy time should not exceed 7 minutes in childbearing-age female patients undergoing barium enema examinations. 6 refs., 1 fig., 2 tabs.