Title: Evaluation of a synthetic single-crystal diamond detector for relative dosimetry on the Leksell Gamma Knife Perfexion radiosurgery system

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4927569· OSTI ID:22581345
; ; ; ; ; ;  [1]; ; ;  [2];  [3]; ;  [4]
  1. Physics Service of Radiation Oncology Department, Clinical and Research Center, Rozzano, Milan 20098 (Italy)
  2. Radiation Oncology Department, Humanitas Clinical and Research Center, Rozzano, Milan 20089 (Italy)
  3. Neurosurgery Department, Humanitas Clinical and Research Center, Rozzano, Milan 20089 (Italy)
  4. Dipartimento di Ingegneria Industriale, Università di Roma Tor Vergata, Roma 00133 (Italy)

Purpose: To evaluate the new commercial PTW-60019 synthetic single-crystal microDiamond detector (PTW, Freiburg, Germany) for relative dosimetry measurements on a clinical Leksell Gamma Knife Perfexion radiosurgery system. Methods: Detector output ratios (DORs) for 4 and 8 mm beams were measured using a microDiamond (PTW-60019), a stereotactic unshielded diode [IBA stereotactic field detector (SFD)], a shielded diode (IBA photon field detector), and GafChromic EBT3 films. Both parallel and transversal acquisition directions were considered for PTW-60019 measurements. Measured DORs were compared to the new output factor reference values for Gamma Knife Perfexion (0.814 and 0.900 for 4 and 8 mm, respectively). Profiles in the three directions were also measured for the 4 mm beam to evaluate full width at half maximum (FWHM) and penumbra and to compare them with the corresponding Leksell GammaPlan profiles. Results: FWHM and penumbra for PTW-60019 differed from the calculated values by less than 0.2 and 0.3 mm, for the parallel and transversal acquisitions, respectively. GafChromic films showed FWHM and penumbra within 0.1 mm. The output ratio obtained with the PTW-60019 for the 4 mm field was 1.6% greater in transverse direction compared to the nominal value. Comparable differences up to 0.8% and 1.0% for, respectively, GafChromic films and SFD were found. Conclusions: The microDiamond PTW-60019 is a suitable detector for commissioning and routine use of Gamma Knife with good agreement of both DORs and profiles in the three directions.

OSTI ID:
22581345
Journal Information:
Medical Physics, Vol. 42, Issue 9; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English