skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SU-E-T-282: Dose Measurements with An End-To-End Audit Phantom for Stereotactic Radiotherapy

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924644· OSTI ID:22548339
;  [1];  [2];  [1]
  1. Calvary Mater Newcastle, Newcastle, NSW (Australia)
  2. University of Sydney, Sydney, NSW (Australia)

Purpose: Report on dose measurements as part of an end-to-end test for stereotactic radiotherapy, using a new audit tool, which allows audits to be performed efficiently either by an onsite team or as a postal audit. Methods: Film measurements have been performed with a new Stereotactic Cube Phantom. The phantom has been designed to perform Winston Lutz type position verification measurements and dose measurements in one setup. It comprises a plastic cube with a high density ball in its centre (used for MV imaging with film or EPID) and low density markers in the periphery (used for Cone Beam Computed Tomography, CBCT imaging). It also features strategically placed gold markers near the posterior and right surfaces, which can be used to calculate phantom rotations on MV images. Slit-like openings allow insertion of film or other detectors.The phantom was scanned and small field treatment plans were created. The fields do not traverse any inhomogeneities of the phantom on their paths to the measurement location. The phantom was setup at the delivery system using CBCT imaging. The calculated treatment fields were delivered, each with a piece of radiochromic film (EBT3) placed in the anterior film holder of the phantom. MU had been selected in planning to achieve similar exposures on all films. Calibration films were exposed in solid water for dose levels around the expected doses. Films were scanned and analysed following established procedures. Results: Setup of the cube showed excellent suitability for CBCT 3D alignment. MV imaging with EPID allowed for clear identification of all markers. Film based dose measurements showed good agreement for MLC created fields down to 0.5 mm × 0.5 mm. Conclusion: An end-to-end audit phantom for stereotactic radiotherapy has been developed and tested.

OSTI ID:
22548339
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English