Title: SU-E-I-79: Source Geometry Dependence of Gamma Well-Counter Measurements

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924076· OSTI ID:22494021
; ;  [1]
  1. Brigham and Women’s Hospital and Harvard Medical School, Boston, MA (United States)

Purpose: To determine the effect of liquid sample volume and geometry on counting efficiency in a gamma well-counter, and to assess the relative contributions of sample geometry and self-attenuation. Gamma wellcounters are standard equipment in clinical and preclinical studies, for measuring patient blood radioactivity and quantifying animal tissue uptake for tracer development and other purposes. Accurate measurements are crucial. Methods: Count rates were measured for aqueous solutions of 99m- Tc at four liquid volume values in a 1-cm-diam tube and at six volume values in a 2.2-cm-diam vial. Total activity was constant for all volumes, and data were corrected for decay. Count rates from a point source in air, supported by a filter paper, were measured at seven heights between 1.3 and 5.7 cm from the bottom of a tube. Results: Sample volume effects were larger for the tube than for the vial. For the tube, count efficiency relative to a 1-cc volume ranged from 1.05 at 0.05 cc to 0.84 at 3 cc. For the vial, relative count efficiency ranged from 1.02 at 0.05 cc to 0.87 at 15 cc. For the point source, count efficiency relative to 1.3 cm from the tube bottom ranged from 0.98 at 1.8 cm to 0.34 at 5.7 cm. The relative efficiency of a 3-cc liquid sample in a tube compared to a 1-cc sample is 0.84; the average relative efficiency for the solid sample in air between heights in the tube corresponding to the surfaces of those volumes (1.3 and 4.8 cm) is 0.81, implying that the major contribution to efficiency loss is geometry, rather than attenuation. Conclusion: Volume-dependent correction factors should be used for accurate quantitation radioactive of liquid samples. Solid samples should be positioned at the bottom of the tube for maximum count efficiency.

OSTI ID:
22494021
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

A COMPARISON OF MEASURED AND CALCULATED GAMMA RAY ATTENUATION FOR A COMMON COUNTING GEOMETRY
Technical Report · Thu Feb 26 00:00:00 EST 2004 · OSTI ID:22494021

Low Level Calibration of Liquid Scintillation Counting
Conference · Wed Jul 01 00:00:00 EDT 2020 · OSTI ID:22494021

Radioassay of yttrium-90 radiation using the radionuclide dose calibrator
Journal Article · Thu May 01 00:00:00 EDT 1997 · Journal of Nuclear Medicine · OSTI ID:22494021