skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: A new development on measurement and control software of SANS BATAN spectrometer (SMARTer) in Serpong, Indonesia

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.4917111· OSTI ID:22391484
; ;  [1]
  1. Center for Technology of Nuclear Industrial Materials, National Nuclear Energy Agency of Indonesia (BATAN) Kawasan Puspiptek Serpong, Tangerang 15314, Banten, Indonesia bharoto@batan.go.id (Indonesia)

In 2005, the main computer for data acquisition and control system of Small-angle Neutron Scattering (SANS) BATAN Spectrometer (SMARTer) was replaced since it halted to operate the spectrometer. According to this replacement, the new software for data acquisition and control system has been developed in-house. Visual Basic programming language is used in developing the software. In the last two years, many developments have been made both in the hardware and also the software to conduct the experiment is more effective and efficient. Lately, the previous motor controller card (ISA Card) was replaced with the programmable motor controller card (PCI Card) for driving one motor of position sensitive detector (PSD), eight motors of four collimators, and six motors of six pinhole discs. This new control system software makes all motors can be moved simultaneously, then it reduces significantly the consuming time of setting up the instrument before running the experiment. Along with that development, the new data acquisition software under MS Windows operating system is also developed to drive a beam stopper in X-Y directions as well as to read the equipment status such as position of the collimators and PSD, to acquire neutron counts on monitor and PSD detectors, and also to manage 12 samples position automatically. A timer object which is set in one second to read the equipment status via serial port of the computer (RS232C), and general purpose interface board (GPIB) for reading the total counts of each pixel of the PSD from histogram memory was used in this new software. The experiment result displayed in real time on the main window, and the data is saved in the special format for further data reduction and analysis. The new software has been implemented and performed for experiment using a preset count or preset time mode for absolute scattering intensity method.

OSTI ID:
22391484
Journal Information:
AIP Conference Proceedings, Vol. 1656, Issue 1; Conference: APS 2012: 5. Asian Physics Symposium, Bandung (Indonesia), 10-12 Jul 2012; Other Information: (c) 2015 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
Country of Publication:
United States
Language:
English