skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Development of multichannel low-energy neutron spectrometer

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4895826· OSTI ID:22303472
; ; ; ; ; ; ; ; ; ; ;  [1];  [2]
  1. Institute of Laser Engineering, Osaka University, 2-6 Yamada-oka, Suita, Osaka (Japan)
  2. Kumamoto University, 2-40-1 Kurokami, Kumamoto 860-8555 (Japan)

A multichannel low-energy neutron spectrometer for down-scattered neutron (DSN) measurements in inertial confinement fusion (ICF) experiments has been developed. Our compact-size 256-channel lithium-glass-scintillator-based spectrometer has been implemented and tested in ICF experiments with the GEKKO XII laser. We have performed time calibration of the 256-channel analog-to-digital convertor system used for DSN measurements via X-ray pulse signals. We have clearly observed the DD-primary fusion neutron signal and have successfully studied the detector's impulse response. Our detector is soon to be implemented in future ICF experiments.

OSTI ID:
22303472
Journal Information:
Review of Scientific Instruments, Vol. 85, Issue 11; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English