Title: Optimization of the x-ray monitoring angle for creating a correlation model between internal and external respiratory signals

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4754648· OSTI ID:22099049
; ; ; ; ; ; ; ; ;  [1]
  1. Department of Radiation Oncology and Image-applied Therapy, Graduate School of Medicine, Kyoto University, Kyoto 606-8507 (Japan)

Purpose: To perform dynamic tumor tracking irradiation with the Vero4DRT (MHI-TM2000), a correlation model [four dimensional (4D) model] between the displacement of infrared markers on the abdominal wall and the three-dimensional position of a tumor indicated by a minimum of three implanted gold markers is required. However, the gold markers cannot be detected successfully on fluoroscopic images under the following situations: (1) overlapping of the gold markers; and (2) a low intensity ratio of the gold marker to its surroundings. In the present study, the authors proposed a method to readily determine the optimal x-ray monitoring angle for creating a 4D model utilizing computed tomography (CT) images. Methods: The Vero4DRT mounting two orthogonal kV x-ray imaging subsystems can separately rotate the gantry along an O-shaped guide-lane and the O-ring along its vertical axis. The optimal x-ray monitoring angle was determined on CT images by minimizing the root-sum-square of water equivalent path lengths (WEPLs) on the orthogonal lines passing all of the gold markers while rotating the O-ring and the gantry. The x-ray monitoring angles at which the distances between the gold markers were within 5 mm at the isocenter level were excluded to prevent false detection of the gold markers in consideration of respiratory motions. First, the relationship between the WEPLs (unit: mm) and the intensity ratios of the gold markers was examined to assess the validity of our proposed method. Second, our proposed method was applied to the 4D-CT images at the end-expiration phase for 11 lung cancer patients who had four to five gold markers. To prove the necessity of the x-ray monitoring angle optimization, the intensity ratios of the least visible markers (minimum intensity ratios) that were estimated from the WEPLs were compared under the following conditions: the optimal x-ray monitoring angle and the angles used for setup verification. Additionally, the intra- and interfractional variations in the intensity ratio were examined from the optimal x-ray monitoring angle. Results: A negative strong correlation was observed between the WEPL (x) and the intensity ratio (y) (y= 6.57 exp[-0.0125x]+ 1, R=-0.88 [95% confidence interval: -0.85 to -0.90], p < 0.01). Our proposed method effectively avoided having the x-ray beam pass through high-density structures, although there were large interpatient variations in the optimal x-ray monitoring angle because of the geometric arrangement between the gold markers and the anatomical structures. The minimum intensity ratios that were estimated from the WEPLs at the optimal x-ray monitoring angle ranged from 1.43 to 2.48, which was an average of 1.27 times (range, 1.02-1.66) higher than the angles used for setup verification. The maximum intra- and interfractional decreases in the intensity ratio were 0.23 and 0.17, respectively. Conclusions: The authors demonstrated that the optimal x-ray monitoring angle for creating a 4D model can improve the visibility of gold markers.

OSTI ID:
22099049
Journal Information:
Medical Physics, Vol. 39, Issue 10; Other Information: (c) 2012 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English