skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Relay NBS Graft with the Plus Delivery System to Improve Deployment in Aortic Arch with Small Radius Curve

Journal Article · · Cardiovascular and Interventional Radiology
; ;  [1]; ;  [2]
  1. San Martino University Hospital, Department of Diagnostic and Interventional Radiology (Italy)
  2. San Martino University Hospital, Department of Cardiac Surgery (Italy)

The purpose of this report is to describe deployment of the Relay NBS Thoracic Stent Graft with the Plus Delivery System (Bolton Medical, Sunrise, FL) in a flexible resin arch model with a 15-mm radius curve as well as our preliminary clinical results. The Relay NBS graft with the Plus Delivery System was evaluated by way of bench testing, which was performed with stent grafts with diameters ranging from 24 to 46 mm and lengths ranging from 100 to 250 mm in flexible resin arch models with a 15-mm arch radius of curvature. The deployment sequence was analyzed. The Relay NBS graft with the Plus Delivery System was deployed in two patients, respectively, having a 6.5-cm penetrating aortic ulcer of the proximal third of the descending thoracic aorta and a DeBakey type-I aortic dissection with chronic false lumen dilatation after surgery due to an entry site at the distal thoracic aorta. Bench tests showed proper conformation and apposition of the Relay NBS graft with the Plus Delivery System in the flexible resin model. This stent graft was deployed successfully into the two patients with a correct orientation of the first stent and without early or late complications. The Relay NBS graft with the Plus Delivery System ensures an optimal conformation and apposition of the first stent in the aortic arch with a small radius of curvature.

OSTI ID:
21608727
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 34, Issue 2; Other Information: DOI: 10.1007/s00270-010-9852-8; Copyright (c) 2011 Springer Science+Business Media, LLC and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE); Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English

Similar Records

Fenestrated Thoracic Endovascular Aortic Repair Using Physician-Modified Stent Grafts (PMSGs) in Zone 0 and Zone 1 for Aortic Arch Diseases
Journal Article · Tue Jan 15 00:00:00 EST 2019 · Cardiovascular and Interventional Radiology · OSTI ID:21608727

Early Results of Chimney Technique for Type B Aortic Dissections Extending to the Aortic Arch
Journal Article · Fri Jan 15 00:00:00 EST 2016 · Cardiovascular and Interventional Radiology · OSTI ID:21608727

Endovascular Repair of Complex Aortic Aneurysms: Intravascular Ultrasound Guidance with an Intracardiac Probe
Journal Article · Mon Sep 15 00:00:00 EDT 2003 · Cardiovascular and Interventional Radiology · OSTI ID:21608727