Title: StarClose Vascular Closure Device: Prospective Study on 222 Deployments in an Interventional Radiology Practice

Journal Article · · Cardiovascular and Interventional Radiology

The StarClose device (Abbott Vascular Devices; Abbott Laboratories, Redwood City, CA) utilizes an externally placed Nitinol clip to achieve arterial closure following femoral artery puncture. The objectives of this study were to assess the efficacy and complications of the StarClose device in patients undergoing interventional radiological procedures. Preprocedural clotting status, pulse and blood pressure, severity of vessel calcification, sheath size, and time to deployment were recorded. Postdeployment complications immediately postprocedure, at 1 h, at 2 h, and at 1 week were recorded. A duplex scan was performed in the first 10 patients to assess any immediate vascular complications. Deployments were successful in 96% achieving immediate hemostasis. Mean deployment time was 48 s. There were no major complications. The StarClose device was found to have a high technical and clinical efficacy.

OSTI ID:
21090839
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 30, Issue 4; Other Information: DOI: 10.1007/s00270-007-9079-5; Copyright (c) 2007 Springer Science+Business Media, LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English