Beam hardening: Analytical considerations of the effective attenuation coefficient of x-ray tomography
- Kramers Laboratorium voor Fysische Technology, Department of Multi-Scale Physics, Delft University of Technology, Pr. Bernhardlaan 6, 2628 Delft (Netherlands)
Polychromatic x-ray beams traveling though material are prone to beam hardening, i.e., the high energy part of the incident spectrum gets over represented when traveling farther into the material. This study discusses the concept of a mean attenuation coefficient in a formal way. The total energy fluence is one-to-one related to the traveled distance in case of a polychromatic beam moving through a given, inhomogeneous material. On the basis of this one-to-one relation, it is useful to define a mean attenuation coefficient and study its decrease with depth. Our results are based on a novel parametrization of the energy dependence of the attenuation coefficient that allows for closed form evaluation of certain spectral integrals. This approach underpins the ad hoc semianalytical expressions given in the literature. An analytical model for the average attenuation coefficient is proposed that uses a simple fit of the attenuation coefficient as a function of the photon energy as input. It is shown that a simple extension of this model gives a rather good description of beam hardening for x-rays traveling through water.
- OSTI ID:
- 20953338
- Journal Information:
- Medical Physics, Vol. 34, Issue 7; Other Information: DOI: 10.1118/1.2742501; (c) 2007 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
Beam Hardening Corrections in Quantitative Computed Tomography
X-ray computed tomography for structural ceramic applications: Beam hardening corrections