Computer tomography (CT) scanners with an increasing number of detector rows offer the potential of shorter scanning times. Nevertheless, the reconstruction problem becomes more challenging, since cone beam artifacts are likely to enter. Here, we consider helical cardiac CT. We analyze how a relationship can be established between exact reconstruction algorithms and the demand to perform a cardiac gating. Utilizing the redundancies requires the consideration of all kinds of Radon planes. For the reconstruction algorithm proposed here, we separate the data into two parts. The first part contains contributions of Radon planes, which are measured with a large number of redundancies. The second part contains the remaining contributions. As it turns out, the second part contributes rather to the low-frequency contents of trans-axial slices. Therefore, we propose to perform a gated back-projection only for the first part, while the second part is back-projected in an ungated way. Data from the complete source trajectory are employed in the reconstruction process in contrary to conventional helical cardiac reconstruction methods. Moreover, all different types of Radon planes are taken into account in the reconstruction, though an ECG-dependent cardiac gating is applied. The reconstruction results, which we present for clinical and simulated data, demonstrate the high potential of CEnPiT for helical cardiac CT with large cone angle systems.

OSTI ID:
20853379
Journal Information:
Medical Physics, Vol. 33, Issue 8; Other Information: DOI: 10.1118/1.2219776; (c) 2006 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

MO-PIS-Exhibit Hall-01: Imaging: CT Dose Optimization Technologies I
Journal Article · Sun Jun 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:20853379

Cardiac cone-beam CT volume reconstruction using ART
Journal Article · Fri Apr 01 00:00:00 EST 2005 · Medical Physics · OSTI ID:20853379

A motion-compensated scheme for helical cone-beam reconstruction in cardiac CT angiography
Journal Article · Tue Jul 15 00:00:00 EDT 2008 · Medical Physics · OSTI ID:20853379