Title: A permanent breast seed implant as partial breast radiation therapy for early-stage patients: A comparison of palladium-103 and iodine-125 isotopes based on radiation safety considerations

Journal Article · · International Journal of Radiation Oncology, Biology and Physics
 [1];  [1];  [2];  [1];  [2]
  1. Department of Medical Physics, Sunnybrook and Women's Health Sciences Centre, Toronto, Ontario (Canada)
  2. Department of Radiation Oncology, Sunnybrook and Women's Health Sciences Centre, Toronto, Ontario (Canada)

Purpose: A permanent breast seed implant (PBSI) technique has been developed as a new form of partial adjuvant radiation therapy for early-stage breast cancer. This study compares iodine-125 ({sup 125}I) and palladium-103 ({sup 103}Pd) isotopes by examining the exposure and effective dose (ED) to a patient's partner.Methods and Materials: A low-energy survey meter was used to measure exposure rates as a function of bolus thickness placed over {sup 103}Pd or {sup 125}I seeds. A general mathematical expression for the initial exposure rate at 1 m (x{sub o,1m}) from the skin surface as a function of the implant size, R, and the distance between the skin surface and the implant, d, was derived. Also, a second general equation is proposed to calculate the ED to the patient's partner.Results: The initial exposure rate at 1 meter and the ED are calculated as follows: x{sub o,1m} = (3{alpha})/2R{sup 3}{center_dot}{beta}{sup 3} [e{sup -{beta}}{sup (2R+d)}({beta}R + 1) + e{sup -{beta}}{sup {center_dot}}{sup d}({beta}R - 1)], and ED = aR{sup b} {center_dot} [e{sup -c(2R+d)} {center_dot} (cR + 1) + e{sup -cd} -bar (cR - 1)]. For {sup 125}I, the parameters are: {alpha} = 0.154409, {beta} = 0.388460, a = 197, b = -0.95, and c = 0.38846. For {sup 103}Pd, they are: {alpha} = 0.06877, {beta} = 0.421098, a = 18.6, b -0.78, and c = 0.421098. For implant diameters varying from 2 to 6 cm and skin-to-implant distances varying from 0.7 to 4 cm, the ED is consistently below 2.6 mSv using the {sup 103}Pd isotope, but more than 5 mSv in many instances and possibly up to 20 mSv using {sup 125}I.Conclusions: PBSI using {sup 103}Pd seeds appears safe because the patient's partner ED is consistently below 5 mSv. The{sup 125}I isotope is not recommended for PBSI.

OSTI ID:
20698447
Journal Information:
International Journal of Radiation Oncology, Biology and Physics, Vol. 62, Issue 2; Other Information: DOI: 10.1016/j.ijrobp.2004.10.014; PII: S0360-3016(04)02787-7; Copyright (c) 2005 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved; Country of input: International Atomic Energy Agency (IAEA); ISSN 0360-3016
Country of Publication:
United States
Language:
English

Similar Records

A Radiation Badge Survey for Family Members Living With Patients Treated With a {sup 103}Pd Permanent Breast Seed Implant
Journal Article · Tue Jan 01 00:00:00 EST 2008 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:20698447

Tolerance and Acceptance Results of a Palladium-103 Permanent Breast Seed Implant Phase I/II Study
Journal Article · Wed Apr 01 00:00:00 EDT 2009 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:20698447

First report of a permanent breast {sup 103}Pd seed implant as adjuvant radiation treatment for early-stage breast cancer
Journal Article · Sun Jan 01 00:00:00 EST 2006 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:20698447