skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Near-field dosimetry of {sup 125}I sources for interstitial brachytherapy implants measured using thermoluminescent sheets

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.1820012· OSTI ID:20634532
; ;  [1]
  1. Department of Oncoradiology, Nara Medical University, 840 Shijo-cho Kashihara-city Nara 634 (Japan)

The dosimetric characteristics were measured for two types of {sup 125}I low-energy photon-emitting sources by using a wide and highly sensitive thermoluminescent (TL) sheet film, which was developed for two-dimensional dose distribution measurements. The TL film is made of Teflon homogeneously mixed with small powders of thermoluminescence (BaSO{sub 4}:Eu doped). Various dosimetric parameters (i.e., radial dose function, 2D and 1D anisotropy functions) of model 6711 and 6702 {sup 125}I sources were obtained at various distances from the source surfaces to 15 mm. These parameters obtained with TL sheet were compared with the data recommended in the updated AAPM TG-43 report. The radial dose functions measured with TL sheet are in agreement with those established data of model 6711 {sup 125}I seed and model 6702 {sup 125}I seed at most of the distances within 5% and 7%, respectively. All the measured anisotropy functions showed symmetry about the longitudinal source axis. The anisotropy of dose distributions was clearly present in the immediate vicinity of the source edges. The measured 2D anisotropy function values at 1 cm are in reasonably good agreement with the recommended values. The differences at two points in the 1D anisotropy functions measured with TL sheet and the established data at 1 cm from source center were 0.7% and 1.9% for model 6711 and 6702 {sup 125}I sources, respectively; the differences at 0.5 cm were 1.5% and 1.7% for model 6711 and 6702 {sup 125}I sources, respectively. The relative dosimetric characteristics in the vicinity of actual interstitial brachytherapy sources containing {sup 125}I have been experimentally determined by using the TL sheet as a 2D dosimeter.

OSTI ID:
20634532
Journal Information:
Medical Physics, Vol. 31, Issue 12; Other Information: DOI: 10.1118/1.1820012; (c) 2004 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English