Title: What is the best method to fit time-resolved data? A comparison of the residual minimization and the maximum likelihood techniques as applied to experimental time-correlated, single-photon counting data

Journal Article · · Journal of Physical Chemistry. B, Condensed Matter, Materials, Surfaces, Interfaces and Biophysical Chemistry
 [1];  [2];  [1];  [1];  [2];  [1]
  1. Iowa State Univ., Ames, IA (United States); Ames Lab., Ames, IA (United States)
  2. Iowa State Univ., Ames, IA (United States)

The need for measuring fluorescence lifetimes of species in subdiffraction-limited volumes in, for example, stimulated emission depletion (STED) microscopy, entails the dual challenge of probing a small number of fluorophores and fitting the concomitant sparse data set to the appropriate excited-state decay function. This need has stimulated a further investigation into the relative merits of two fitting techniques commonly referred to as “residual minimization” (RM) and “maximum likelihood” (ML). Fluorescence decays of the well-characterized standard, rose bengal in methanol at room temperature (530 ± 10 ps), were acquired in a set of five experiments in which the total number of “photon counts” was approximately 20, 200, 1000, 3000, and 6000 and there were about 2–200 counts at the maxima of the respective decays. Each set of experiments was repeated 50 times to generate the appropriate statistics. Each of the 250 data sets was analyzed by ML and two different RM methods (differing in the weighting of residuals) using in-house routines and compared with a frequently used commercial RM routine. Convolution with a real instrument response function was always included in the fitting. While RM using Pearson’s weighting of residuals can recover the correct mean result with a total number of counts of 1000 or more, ML distinguishes itself by yielding, in all cases, the same mean lifetime within 2% of the accepted value. For 200 total counts and greater, ML always provides a standard deviation of <10% of the mean lifetime, and even at 20 total counts there is only 20% error in the mean lifetime. Here, the robustness of ML advocates its use for sparse data sets such as those acquired in some subdiffraction-limited microscopies, such as STED, and, more importantly, provides greater motivation for exploiting the time-resolved capacities of this technique to acquire and analyze fluorescence lifetime data.

Research Organization:
Ames Laboratory (AMES), Ames, IA (United States)
Sponsoring Organization:
USDOE Office of Science (SC), Basic Energy Sciences (BES)
Grant/Contract Number:
CCF-1117125; W-7405-430 ENG-82; AC02-07CH11358
OSTI ID:
1240907
Report Number(s):
IS-J-8935
Journal Information:
Journal of Physical Chemistry. B, Condensed Matter, Materials, Surfaces, Interfaces and Biophysical Chemistry, Vol. 120, Issue 9; ISSN 1520-6106
Publisher:
American Chemical SocietyCopyright Statement
Country of Publication:
United States
Language:
English
Citation Metrics:
Cited by: 19 works
Citation information provided by
Web of Science

References (37)

Phasor approach to fluorescence lifetime microscopy distinguishes different metabolic states of germ cells in a live tissue journal August 2011
A novel fluorescence lifetime imaging system that optimizes photon efficiency journal January 2008
Supercontinuum Stimulated Emission Depletion Fluorescence Lifetime Imaging journal June 2012
The Number of Accumulated Photons and the Quality of Stimulated Emission Depletion Lifetime Images journal March 2014
How many photons are necessary for fluorescence-lifetime measurements? journal November 1992
Fluorescence Instrument Response Standards in Two-Photon Time-Resolved Spectroscopy journal August 2010
Towards multi-exponential analysis in optically stimulated luminescence journal April 2010
Maximum likelihood method for the analysis of time-resolved fluorescence decay curves journal December 1991
Clarification of the use of CHI-square and likelihood functions in fits to histograms journal April 1984
On the analysis of fluorescence decay kinetics by the method of least-squares journal June 1974
Better estimates of exponential decay parameters journal October 1981
Comparison of maximum likelihood estimation and chi-square statistics applied to counting experiments journal January 2001
On the reliability of fluorescence decay data journal January 1977
Single-Molecule Analysis And Lifetime Estimates Of Heterogeneous Low-Count-Rate Time-Correlated Fluorescence Data journal September 2011
Efficient maximum likelihood estimator fitting of histograms journal May 2010
An Experimental Comparison of the Maximum Likelihood Estimation and Nonlinear Least-Squares Fluorescence Lifetime Analysis of Single Molecules journal May 2001
Comparison of Methods for Rapid Evaluation of Lifetimes of Exponential Decays journal May 2004
Artefacts in the analysis of temporal response functions measured by photon counting journal March 2005
Analysis of Fluorescence Decay by the Nonlinear Least Squares Method journal November 1988
Error Analysis of the Rapid Lifetime Determination Method for Double-Exponential Decays and New Windowing Schemes journal March 1999
Accurate Analysis of Fluorescence Decays from Single Molecules in Photon Counting Experiments journal August 2003
Analysis of fluorescence lifetime data for single Rhodamine molecules in flowing sample streams journal January 1994
Comparison between time-correlated single photon counting and fluorescence correlation spectroscopy in single molecule identification journal March 1998
Data analysis in the shot noise limit. 1. Single parameter estimation with Poisson and normal probability density functions journal November 1989
Data analysis in the shot noise limit. 2. Methods for data regression journal November 1989
On the Use and Interpretation of Certain Test Criteria for Purposes of Statistical Inference: Part I journal July 1928
On the Use and Interpretation of Certain Test Criteria for Purposes of Statistical Inference: Part II journal December 1928
A Bayesian method for single molecule, fluorescence burst analysis journal January 2010
Bayesian analysis of fluorescence lifetime imaging data conference February 2011
Ultrafast method for the analysis of fluorescence lifetime imaging microscopy data based on the Laguerre expansion technique journal January 2005
Reconvolution analysis in time‐resolved fluorescence experiments—an alternative approach: Reference‐to‐excitation‐to‐fluorescence reconvolution journal December 1993
Exponential analysis in physical phenomena journal February 1999
Wavelength dependence of timing properties of the XP 2020 photomultiplier journal December 1978
Wavelength dependence of the time resolution of high-speed photomultipliers used in single-photon timing experiments journal January 1976
Systematic errors in χ2-fitting of Poisson distributions journal March 1996
The Likelihood Test of Independence in Contingency Tables journal December 1935
The Large-Sample Distribution of the Likelihood Ratio for Testing Composite Hypotheses journal March 1938

Cited By (3)

Giant light-harvesting nanoantenna for single-molecule detection in ambient light journal September 2017
Unveiling the Photo‐ and Thermal‐Stability of Cesium Lead Halide Perovskite Nanocrystals journal October 2019
A Bayesian Approach for Extracting Fluorescence Lifetimes from Sparse Data Sets and Its Significance for Imaging Experiments journal December 2018

Similar Records

Photon Counting Data Analysis: Application of the Maximum Likelihood and Related Methods for the Determination of Lifetimes in Mixtures of Rose Bengal and Rhodamine B
Journal Article · Mon Dec 12 00:00:00 EST 2016 · Journal of Physical Chemistry. A, Molecules, Spectroscopy, Kinetics, Environment, and General Theory · OSTI ID:1240907

A Bayesian Approach for Extracting Fluorescence Lifetimes from Sparse Data Sets and Its Significance for Imaging Experiments
Journal Article · Sat Dec 22 00:00:00 EST 2018 · Photochemistry and Photobiology · OSTI ID:1240907

The Number of Accumulated Photons and the Quality of Stimulated Emission Depletion Lifetime Images
Journal Article · Mon Mar 03 00:00:00 EST 2014 · Photochemistry and Photobiology · OSTI ID:1240907