Title: Performance of microchannel plate based detectors for <25 keV x rays: Monte Carlo simulations and comparisons with experimental results

Abstract

We present the results of Monte Carlo simulations of the microchannel plate (MCP) response to x rays in the 250 eV to 25 keV energy range as a function of both x-ray energy and impact angle and their comparisons with the experimental results from the X8A beamline at the National Synchrotron Light Source at Brookhaven National Laboratory. Incoming x rays interact with the lead glass of the microchannel plate, producing photoelectrons. Transport of the photoelectrons is neglected in this model, and it is assumed that photoelectrons deposit all their energy at the point they are created. This deposition leads to the generation of many secondary electrons, some fraction of which diffuse to the MCP pore surface where they can initiate secondary electron cascades in the pore under an external voltage bias. X-ray penetration through multiple MCP pore walls is increasingly important above 5 keV, and the effect of this penetration on MCP performance is studied. In agreement with past measurements, we find that the dependence of MCP sensitivity with angle relative to the pore bias changes from a cotangent dependence to angular independence and then proceeds to a secant dependence as the x-ray energy increases. We also find that withmore » the increasing x-ray energy, the MCP gain sensitivity as a function of bias voltage decreases. The simulations also demonstrate that for x rays incident normal to the MCP surface, spatial resolution shows little dependence on the x-ray energy but degrades with the increasing x-ray energy as the angle of incidence relative to the surface normal increases. This agrees with experimental measurements. Simulation studies have also been completed for MCPs gated with a subnanosecond voltage pulse. In this work, we find that the optical gate profile width increases as the x-ray energy is increased above 5 keV, a consequence of increased x-ray penetration at energies >5 keV. Simulations of the pulsed dynamic range show that the dynamic range varies between ~100 and 1000 depending on x-ray energy and peak voltage.« less

Authors:
 [1]; ORCiD logo [2]; ORCiD logo [3]
  1. Nevada National Security Site (NNSS), Los Alamos, NV (United States). New Mexico Operations; Los Alamos National Lab. (LANL), Los Alamos, NM (United States)
  2. Sandia National Lab. (SNL-NM), Albuquerque, NM (United States)
  3. Nevada National Security Site, Santa Barbara, CA (United States). Special Technologies Laboratory
Publication Date:
Research Org.:
Sandia National Lab. (SNL-NM), Albuquerque, NM (United States); Nevada National Security Site/Mission Support and Test Services LLC (NNSS/MSTS), North Las Vegas, NV (United States)
Sponsoring Org.:
USDOE National Nuclear Security Administration (NNSA), Office of Defense Programs (DP)
OSTI Identifier:
1781537
Alternate Identifier(s):
OSTI ID: 1773562
Report Number(s):
SAND-2020-10815J; DOE/NV/03624-0813
Journal ID: ISSN 0034-6748; 691312; TRN: US2210299
Grant/Contract Number:  
AC04-94AL85000; NA0003624; NA0003525; DENA-0003525
Resource Type:
Accepted Manuscript
Journal Name:
Review of Scientific Instruments
Additional Journal Information:
Journal Volume: 92; Journal Issue: 4; Journal ID: ISSN 0034-6748
Publisher:
American Institute of Physics (AIP)
Country of Publication:
United States
Language:
English
Subject:
47 OTHER INSTRUMENTATION; Monte Carlo methods; Synchrotron radiation; High-energy X rays; Quantum efficiency; Computer simulation; Microchannel plate detectors

Citation Formats

Kruschwitz, Craig A., Wu, Ming, and Moy, Ken. Performance of microchannel plate based detectors for <25 keV x rays: Monte Carlo simulations and comparisons with experimental results. United States: N. p., 2021. Web. doi:10.1063/5.0030514.
Kruschwitz, Craig A., Wu, Ming, & Moy, Ken. Performance of microchannel plate based detectors for <25 keV x rays: Monte Carlo simulations and comparisons with experimental results. United States. https://doi.org/10.1063/5.0030514
Kruschwitz, Craig A., Wu, Ming, and Moy, Ken. Thu . "Performance of microchannel plate based detectors for <25 keV x rays: Monte Carlo simulations and comparisons with experimental results". United States. https://doi.org/10.1063/5.0030514. https://www.osti.gov/servlets/purl/1781537.
@article{osti_1781537,
title = {Performance of microchannel plate based detectors for <25 keV x rays: Monte Carlo simulations and comparisons with experimental results},
author = {Kruschwitz, Craig A. and Wu, Ming and Moy, Ken},
abstractNote = {We present the results of Monte Carlo simulations of the microchannel plate (MCP) response to x rays in the 250 eV to 25 keV energy range as a function of both x-ray energy and impact angle and their comparisons with the experimental results from the X8A beamline at the National Synchrotron Light Source at Brookhaven National Laboratory. Incoming x rays interact with the lead glass of the microchannel plate, producing photoelectrons. Transport of the photoelectrons is neglected in this model, and it is assumed that photoelectrons deposit all their energy at the point they are created. This deposition leads to the generation of many secondary electrons, some fraction of which diffuse to the MCP pore surface where they can initiate secondary electron cascades in the pore under an external voltage bias. X-ray penetration through multiple MCP pore walls is increasingly important above 5 keV, and the effect of this penetration on MCP performance is studied. In agreement with past measurements, we find that the dependence of MCP sensitivity with angle relative to the pore bias changes from a cotangent dependence to angular independence and then proceeds to a secant dependence as the x-ray energy increases. We also find that with the increasing x-ray energy, the MCP gain sensitivity as a function of bias voltage decreases. The simulations also demonstrate that for x rays incident normal to the MCP surface, spatial resolution shows little dependence on the x-ray energy but degrades with the increasing x-ray energy as the angle of incidence relative to the surface normal increases. This agrees with experimental measurements. Simulation studies have also been completed for MCPs gated with a subnanosecond voltage pulse. In this work, we find that the optical gate profile width increases as the x-ray energy is increased above 5 keV, a consequence of increased x-ray penetration at energies >5 keV. Simulations of the pulsed dynamic range show that the dynamic range varies between ~100 and 1000 depending on x-ray energy and peak voltage.},
doi = {10.1063/5.0030514},
journal = {Review of Scientific Instruments},
number = 4,
volume = 92,
place = {United States},
year = {Thu Apr 01 00:00:00 EDT 2021},
month = {Thu Apr 01 00:00:00 EDT 2021}
}

Works referenced in this record:

Gain model for microchannel plates
journal, January 1979


X‐ray detection characteristics of microchannel plates in the energy range from 1.8 to 8 keV
journal, March 1989

  • Yamaguchi, N.; Cho, T.; Kondoh, T.
  • Review of Scientific Instruments, Vol. 60, Issue 3
  • DOI: 10.1063/1.1140440

Monte Carlo simulations for tilted-channel electron multipliers
journal, June 2000

  • Choi, Y. S.; Kim, J. M.
  • IEEE Transactions on Electron Devices, Vol. 47, Issue 6
  • DOI: 10.1109/16.842976

The characterization of x‐ray photocathodes in the 0.1–10‐keV photon energy region
journal, March 1981

  • Henke, B. . L.; Knauer, J. P.; Premaratne, K.
  • Journal of Applied Physics, Vol. 52, Issue 3
  • DOI: 10.1063/1.329789

Improved modeling of microchannel plate response to hard X-rays
journal, March 2013

  • Farley, D. R.; Izumi, N.; Landen, O. L.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 705
  • DOI: 10.1016/j.nima.2012.12.078

The characterisation of soft X-ray photocathodes in the wavelength band 1–300 Å
journal, February 1983


Measurement and modeling of pulsed microchannel plate operation (invited)
journal, October 2008

  • Rochau, G. A.; Wu, M.; Kruschwitz, C.
  • Review of Scientific Instruments, Vol. 79, Issue 10
  • DOI: 10.1063/1.2965787

Static and time-resolved 10–1000 keV x-ray imaging detector options for NIF
journal, October 2004

  • Landen, O. L.; Bell, P. M.; McDonald, J. W.
  • Review of Scientific Instruments, Vol. 75, Issue 10
  • DOI: 10.1063/1.1787904

Characterization of microchannel plates for plasma x‐ray diagnostics in the 0.6–82‐keV energy region
journal, July 1989

  • Yamaguchi, N.; Cho, T.; Kondoh, T.
  • Review of Scientific Instruments, Vol. 60, Issue 7
  • DOI: 10.1063/1.1140748

Secondary electron emission properties
journal, January 1996


X-Ray Interactions: Photoabsorption, Scattering, Transmission, and Reflection at E = 50-30,000 eV, Z = 1-92
journal, July 1993

  • Henke, B. L.; Gullikson, E. M.; Davis, J. C.
  • Atomic Data and Nuclear Data Tables, Vol. 54, Issue 2, p. 181-342
  • DOI: 10.1006/adnd.1993.1013

Generalized hard x-ray detection model for microchannel plate detectors
journal, October 1997

  • Shikhaliev, P. M.
  • Review of Scientific Instruments, Vol. 68, Issue 10
  • DOI: 10.1063/1.1148011

Angular sensitivity of gated microchannel plate framing cameras
journal, January 2001

  • Landen, O. L.; Lobban, A.; Tutt, T.
  • Review of Scientific Instruments, Vol. 72, Issue 1
  • DOI: 10.1063/1.1318256

Computer Analysis of the Timing Properties in Micro Channel Plate Photomultiplier Tubes
journal, January 1984


Hard X-ray detection model for microchannel plate detectors
journal, October 1997

  • Shikhaliev, P. M.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 398, Issue 2-3
  • DOI: 10.1016/s0168-9002(97)00685-2

Current response characteristics of microchannel plates x‐ray detector using synchrotron radiation (0.6–2 keV and 5–20 keV)
journal, February 1988

  • Kondoh, T.; Yamaguchi, N.; Cho, T.
  • Review of Scientific Instruments, Vol. 59, Issue 2
  • DOI: 10.1063/1.1140236

Monte Carlo simulations of microchannel plate detectors. I. Steady-state voltage bias results
journal, July 2008

  • Wu, Ming; Kruschwitz, Craig A.; Morgan, Dane V.
  • Review of Scientific Instruments, Vol. 79, Issue 7
  • DOI: 10.1063/1.2949119

Characterizations of MCP performance in the hard x-ray range (6–25 keV)
journal, November 2014

  • Wu, Ming; Moy, Ken; Kruschwitz, Craig
  • Review of Scientific Instruments, Vol. 85, Issue 11
  • DOI: 10.1063/1.4890293

Monte Carlo simulations of microchannel plate detectors. II. Pulsed voltage results
journal, February 2011

  • Kruschwitz, Craig A.; Wu, Ming; Rochau, Greg A.
  • Review of Scientific Instruments, Vol. 82, Issue 2
  • DOI: 10.1063/1.3530451

X-ray detection characteristics of gold photocathodes and microchannel plates using synchrotron radiation (10 eV–82.5 keV)
journal, May 1992

  • Hirata, M.; Cho, T.; Takahashi, E.
  • Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, Vol. 66, Issue 4
  • DOI: 10.1016/0168-583x(92)95422-n

Microchannel plate detectors
journal, June 1979


Energy dependent sensitivity of microchannel plate detectors
journal, October 2006

  • Rochau, G. A.; Bailey, J. E.; Chandler, G. A.
  • Review of Scientific Instruments, Vol. 77, Issue 10
  • DOI: 10.1063/1.2336461

Monte Carlo simulations of high-speed, time-gated microchannel-plate-based x-ray detectors: Saturation effects in dc and pulsed modes and detector dynamic range
journal, October 2008

  • Kruschwitz, Craig A.; Wu, Ming; Moy, Ken
  • Review of Scientific Instruments, Vol. 79, Issue 10
  • DOI: 10.1063/1.2969283

Three-dimensional model of x-ray induced microchannel plate output
journal, October 2006

  • Harding, E. C.; Drake, R. P.
  • Review of Scientific Instruments, Vol. 77, Issue 10
  • DOI: 10.1063/1.2220072

Soft-x-ray-induced secondary-electron emission from semiconductors and insulators: Models and measurements
journal, March 1979


Time- and space-resolved elliptical crystal spectrometers for high energy density physics research
journal, October 2004

  • Lake, P. W.; Bailey, J. E.; Rochau, G. A.
  • Review of Scientific Instruments, Vol. 75, Issue 10
  • DOI: 10.1063/1.1788866