Title: Digital pulse processing for NaI(Tl) detectors

Abstract

We apply two different post-processing techniques to digital pulses induced by photons in a NaI(Tl) detector and compare the obtained energy resolution to the standard analog approach. Our digital acquisition approach is performed using a single-stage acquisition with a fast digitizer. Both the post-processing techniques we propose rely on signal integration. In the first, the pulse integral is calculated by directly numerically integrating the pulse digital samples, while in the second the pulse integral is estimated by a model-based fitting of the pulse. Our study used a 7.62 cm×7.62 cm cylindrical NaI(Tl) detector that gave a 7.60% energy resolution (at 662 keV), using the standard analog acquisition approach, based on a pulse shaping amplifier. The new direct numerical integration yielded a 6.52% energy resolution. The fitting approach yielded a 6.55% energy resolution, and, although computationally heavier than numerical integration, is preferable when only the early samples of the pulse are available. We also evaluated the timing performance of a fast–slow detection system, encompassing an EJ-309 and a NaI(Tl) scintillator. Here, we use two techniques to determine the pulse start time: constant fraction discrimination (CFD) and adaptive noise threshold timing (ANT), for both the analog and digital acquisition approach. With themore » analog acquisition approach, we found a system time resolution of 5.8 ns and 7.3 ns, using the constant fraction discrimination and adaptive noise threshold timing, respectively. With the digital acquisition approach, a time resolution of 1.2 ns was achieved using the ANT method and 3.3 ns using CFD at 50% of the maximum, to select the pulse start time. Finally, the proposed direct digital readout and post-processing techniques can improve the application of NaI(Tl) detectors, traditionally considered ‘slow’, for fast counting and correlation measurements, while maintaining a good measurement of the energy resolution.« less

Authors:
 [1];  [1];  [1];  [1]
  1. Univ. of Michigan, Ann Arbor, MI (United States)
Publication Date:
Research Org.:
Univ. of Michigan, Ann Arbor, MI (United States)
Sponsoring Org.:
USDOE National Nuclear Security Administration (NNSA), Office of Nonproliferation and Verification Research and Development (NA-22); USDOE National Nuclear Security Administration (NNSA)
OSTI Identifier:
1365802
Alternate Identifier(s):
OSTI ID: 1397504
Grant/Contract Number:  
NA0002534
Resource Type:
Accepted Manuscript
Journal Name:
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Additional Journal Information:
Journal Volume: 806; Journal Issue: C; Journal ID: ISSN 0168-9002
Publisher:
Elsevier
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; NaI(Tl) detector; Digital acquisition; Data processing; Energy resolution; Time resolution

Citation Formats

Di Fulvio, A., Shin, T. H., Hamel, M. C., and Pozzi, S. A. Digital pulse processing for NaI(Tl) detectors. United States: N. p., 2015. Web. doi:10.1016/j.nima.2015.09.080.
Di Fulvio, A., Shin, T. H., Hamel, M. C., & Pozzi, S. A. Digital pulse processing for NaI(Tl) detectors. United States. https://doi.org/10.1016/j.nima.2015.09.080
Di Fulvio, A., Shin, T. H., Hamel, M. C., and Pozzi, S. A. Fri . "Digital pulse processing for NaI(Tl) detectors". United States. https://doi.org/10.1016/j.nima.2015.09.080. https://www.osti.gov/servlets/purl/1365802.
@article{osti_1365802,
title = {Digital pulse processing for NaI(Tl) detectors},
author = {Di Fulvio, A. and Shin, T. H. and Hamel, M. C. and Pozzi, S. A.},
abstractNote = {We apply two different post-processing techniques to digital pulses induced by photons in a NaI(Tl) detector and compare the obtained energy resolution to the standard analog approach. Our digital acquisition approach is performed using a single-stage acquisition with a fast digitizer. Both the post-processing techniques we propose rely on signal integration. In the first, the pulse integral is calculated by directly numerically integrating the pulse digital samples, while in the second the pulse integral is estimated by a model-based fitting of the pulse. Our study used a 7.62 cm×7.62 cm cylindrical NaI(Tl) detector that gave a 7.60% energy resolution (at 662 keV), using the standard analog acquisition approach, based on a pulse shaping amplifier. The new direct numerical integration yielded a 6.52% energy resolution. The fitting approach yielded a 6.55% energy resolution, and, although computationally heavier than numerical integration, is preferable when only the early samples of the pulse are available. We also evaluated the timing performance of a fast–slow detection system, encompassing an EJ-309 and a NaI(Tl) scintillator. Here, we use two techniques to determine the pulse start time: constant fraction discrimination (CFD) and adaptive noise threshold timing (ANT), for both the analog and digital acquisition approach. With the analog acquisition approach, we found a system time resolution of 5.8 ns and 7.3 ns, using the constant fraction discrimination and adaptive noise threshold timing, respectively. With the digital acquisition approach, a time resolution of 1.2 ns was achieved using the ANT method and 3.3 ns using CFD at 50% of the maximum, to select the pulse start time. Finally, the proposed direct digital readout and post-processing techniques can improve the application of NaI(Tl) detectors, traditionally considered ‘slow’, for fast counting and correlation measurements, while maintaining a good measurement of the energy resolution.},
doi = {10.1016/j.nima.2015.09.080},
journal = {Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment},
number = C,
volume = 806,
place = {United States},
year = {Fri Oct 02 00:00:00 EDT 2015},
month = {Fri Oct 02 00:00:00 EDT 2015}
}

Journal Article:

Citation Metrics:
Cited by: 12 works
Citation information provided by
Web of Science

Save / Share:

Works referenced in this record:

Thallium Halide Crystal Counter
journal, December 1947


Performance evaluation of A-SPECT: a high resolution desktop pinhole SPECT system for imaging small animals
journal, October 2002

  • McElroy, D. P.; MacDonald, L. R.; Beekman, F. J.
  • IEEE Transactions on Nuclear Science, Vol. 49, Issue 5
  • DOI: 10.1109/TNS.2002.803801

Dual-particle imaging system based on simultaneous detection of photon and neutron collision events
journal, October 2014

  • Poitrasson-Rivière, Alexis; Hamel, Michael C.; Polack, J. Kyle
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 760
  • DOI: 10.1016/j.nima.2014.05.056

AGATA—Advanced GAmma Tracking Array
journal, March 2012

  • Akkoyun, S.; Algora, A.; Alikhani, B.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 668
  • DOI: 10.1016/j.nima.2011.11.081

A gamma–gamma coincidence spectrometric method for rapid characterization of uranium isotopic fingerprints
journal, October 2010

  • Zhang, Weihua; Yi, Jing; Mekarski, Pawel
  • Journal of Radioanalytical and Nuclear Chemistry, Vol. 288, Issue 1
  • DOI: 10.1007/s10967-010-0868-y

A new timing model for calculating the intrinsic timing resolution of a scintillator detector
journal, January 2007


Status of timing with plastic scintillation detectors
journal, January 1979


Fundamental limits of scintillation detector timing precision
journal, May 2014

  • Derenzo, Stephen E.; Choong, Woon-Seng; Moses, William W.
  • Physics in Medicine and Biology, Vol. 59, Issue 13
  • DOI: 10.1088/0031-9155/59/13/3261

The nonlinear light output of NaI(Tl) detectors in the Modular Total Absorption Spectrometer
journal, July 2015

  • Rasco, B. C.; Fijałkowska, A.; Karny, M.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 788
  • DOI: 10.1016/j.nima.2015.03.087

Study on a digital pulse processing algorithm based on template-matching for high-throughput spectroscopy
journal, June 2015

  • Wen, Xianfei; Yang, Haori
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 784
  • DOI: 10.1016/j.nima.2014.11.008

Works referencing / citing this record:

Applications in energy spectrum measurement based on pulse fitting
journal, March 2019

  • Liang, Yong-Fei; Yang, Chao-Wen; Xu, Jia-Yun
  • Radiation Detection Technology and Methods, Vol. 3, Issue 3
  • DOI: 10.1007/s41605-019-0096-0