Design and Fabrication of a PDMS Microchip Based Immunoassay

Conference ·
DOI:https://doi.org/10.1117/12.847138· OSTI ID:994039

In this paper, we describe the design and fabrication process of a polydimethylsiloxane (PDMS) microchip for on-chip multiplex immunoassay application. The microchip consists of a PDMS microfluidic channel layer and a micro pneumatic valve control layer. By selectively pressurizing the pneumatic microvalves, immuno reagents were controlled to flow and react in certain fluidic channel sites. Cross contamination was prevented by tightly closed valves. Our design was proposed to utilize PDMS micro channel surface as the solid phase immunoassay substrate and simultaneously detect four targets antigens on chip. Experiment result shows that 20psi valve pressure is sufficient to tightly close a 200µm wide micro channel with flow rate up to 20µl/min.

Research Organization:
Pacific Northwest National Laboratory (PNNL), Richland, WA (US)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-76RL01830
OSTI ID:
994039
Report Number(s):
PNNL-SA-73305
Country of Publication:
United States
Language:
English

Similar Records

Poly(dimethylsiloxane) microchip-based immunoassay with multiple reaction zones: Toward on-chip multiplex detection platform
Journal Article · Tue Sep 20 00:00:00 EDT 2011 · Sensors and Actuators. B, Chemical, 159(1):44-50 · OSTI ID:1026606

Electrokinetic Sample Preconcentration and Hydrodynamic Sample Injection for Microchip Electrophoresis Using a Pneumatic Microvalve
Journal Article · Sun Jan 31 23:00:00 EST 2016 · Electrophoresis · OSTI ID:1243169

Microfabricated Renewable Beads-Trapping/Releasing Flow Cell for Rapid Antigen-Antibody Reaction in Chemiluminescent Immunoassay
Journal Article · Fri Apr 01 00:00:00 EDT 2011 · Analytical Chemistry, 83(7):2685-2690 · OSTI ID:1011784