Title: Report of multijunction thermal convertor interlaboratory comparison study for measurements at Sandia National Laboratory.

Technical Report ·
DOI:https://doi.org/10.2172/985497· OSTI ID:985497

Two Multijunction Thermal Voltage Converters (MJTCs) were provided to the Sandia National Laboratory Primary Standards Laboratory (Sandia PSL) as part of an interlaboratory comparison (ILC). This report summarizes the results of the measurements of the devices (S 127D1 and S 127C2) measured at Sandia PSL from March 4 to March 15, 2009. The SNL/NM portion of an interlaboratory comparison of multijunction thermal convertors was successfully completed with a demonstrated measurement uncertainty of 60ppm (k=2).

Research Organization:
Sandia National Laboratories (SNL), Albuquerque, NM, and Livermore, CA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC04-94AL85000
OSTI ID:
985497
Report Number(s):
SAND2009-4334; TRN: US201016%%2123
Country of Publication:
United States
Language:
English