Experimental heat leak measurements on the APT 210 KW CW RF power coupler

Conference ·
OSTI ID:975635
A cryogenic test rig was designed and fabricated to measure the heat leak from room temperature to 2 K from the Accelerator Production of Tritium (APT) 700 MHz, 210 kW continuous wave (CW) co-axial power coupler (PC). The outer conductor of the PC is stainless steel with 15 {micro}m copper film on the inside. The copper inner conductor operates at room temperature and contributes considerable infra-red radiation heat load to 2 K. Two thermal intercept heat exchangers cooled by supercritical helium are incorporated into the outer conductor to reduce the heat conducted to the lowest temperatures. A brief description of the experimental apparatus is presented. A comparison of the experimental measurements and the predictions of a detailed thermal model is given. There is also a discussion of anomalous behavior observed in the thermal intercepts, and fluctuations in the helium coolant properties.