An instrument design for the accurate determination of the electron beam location in the linac coherant light source undulator.
Conference
·
OSTI ID:972598
The linac coherent light source (LCLS), currently under design, requires accurate alignment between the electron beam and each undulator's magnetic centerline. A beam finder wire (BFW) instrument has been developed to provide beam location information that is used to move the undulators to their appropriate positions. A BFW instrument is mounted at each of the 33 magnets in the undulator section. Beam detection is achieved by electrons impacting two carbon fiber wires and then sensing the downstream radiation. The wires are mounted vertically and horizontally on a wire card similar to that of a traditional wire scanner instrument. The development of the BFW presents several design challenges due to the need for high accuracy of the wires locations and the need for removal of the wires during actual operation of the LCLS (30 microns repeatability is required for the wire locations). In this paper, we present the technical specification, design criteria, mechanical design, and results from prototype tests for the BFW.
- Research Organization:
- Argonne National Laboratory (ANL)
- Sponsoring Organization:
- SC
- DOE Contract Number:
- AC02-06CH11357
- OSTI ID:
- 972598
- Report Number(s):
- ANL/ASD/CP-59591
- Country of Publication:
- United States
- Language:
- ENGLISH
Similar Records
An Instrument Design for the Accurate Determination of the Electron Beam Location in the Linac Coherent Light Source Undulator
A wire scanner design for electron beam profile measurements in the Linac Coherent Light Source undulator.
Electron Signal Detection for the Beam-Finder Wire of the Linac Coherent Light Source Undulator
Conference
·
Tue Jan 22 23:00:00 EST 2008
· Conf.Proc.C070625:4324,2007
·
OSTI ID:922593
A wire scanner design for electron beam profile measurements in the Linac Coherent Light Source undulator.
Conference
·
Fri Dec 31 23:00:00 EST 2004
·
OSTI ID:925243
Electron Signal Detection for the Beam-Finder Wire of the Linac Coherent Light Source Undulator
Conference
·
Thu Sep 21 00:00:00 EDT 2006
·
OSTI ID:891835