A Scanning Electron Microscope Facility for Characterization of Tritium Containing Materials

Conference ·
OSTI ID:971993
A scanning electron microscope (SEM) facility for the examination of tritium-containing materials is operational at Mound Laboratory. The SEM is installed with the sample chamber incorporated as an integral part of an inert gas glovebox facility to enable easy handling of radioactive and pyrophoric materials. A standard SEM (ERTEC Model B-1) was modified to meet dimensional, operational, and safety-related requirements. A glovebox was designed and fabricated which permitted access with the gloves to all parts of the SEM sample chamber to facilitate detector and accessory replacement and repairs. A separate console combining the electron optical column and specimen chamber was interfaced to the glovebox by a custom-made, neoprene bellows so that the vibrations normally associated with the blowers and pumps were damped. Photomicrographs of tritiated pyrophoric materials show the usefulness of this facility. Some of the difficulties involved in the investigation of these materials are also discussed.
Research Organization:
MOUND (Mound Plant, Miamisburg, OH (United States))
Sponsoring Organization:
USDOE
OSTI ID:
971993
Report Number(s):
MLM-2275(OP); CONF-750989 Vol. 4
Country of Publication:
United States
Language:
English

Similar Records

SEM Facility for Examination of Reactive and Radioactive Materials
Technical Report · Fri Jun 01 00:00:00 EDT 1973 · OSTI ID:969922

Monsanto/Mound Laboratory Tritium Effluent Control Systems
Conference · Fri Sep 26 00:00:00 EDT 1975 · OSTI ID:934594

Contained scanning electron microscope facility for examining radioactive materials
Technical Report · Fri Feb 28 23:00:00 EST 1986 · OSTI ID:5493781