Design updates of the x-ray beam position monitor for beamline front ends.
At the Advanced Photon Source (APS), each insertion device (ID) beamline front end has two x-ray beam position monitors (XBPMs) to monitor the x-ray beam position in both the vertical and horizontal directions. The XBPMs measure photoelectrons generated from the CVD-diamond-based sensory blades and deduce the beam position by comparison of the relative signals from the blades. Using the method proposed by G. Decker, which involves the introduction of a chicane into the accelerator lattice that directs unwanted x-rays away from the photosensitive XBPM blades, the photon source stability has been improved by addition of XBPMs in the storage ring global orbit feedback. In recent years, design updates for the XBPM mechanical structure and geometric configuration have been made to improve its performance. We present these design updates in this paper. Test results of the XBPM design updates are also discussed here.
- Research Organization:
- Argonne National Laboratory (ANL)
- Sponsoring Organization:
- SC
- DOE Contract Number:
- AC02-06CH11357
- OSTI ID:
- 967253
- Report Number(s):
- ANL/AOD/CP-117934
- Country of Publication:
- United States
- Language:
- ENGLISH
Similar Records
Design Updates of the X-ray Beam Position Monitor for Beamline Front Ends
Progress of the APS high heat load x-ray beam position monitor development
Progress of the APS high heat load X-ray beam position monitor development
Journal Article
·
Thu Jan 18 23:00:00 EST 2007
· AIP Conference Proceedings
·
OSTI ID:21049190
Progress of the APS high heat load x-ray beam position monitor development
Journal Article
·
Tue Jul 01 00:00:00 EDT 1997
· AIP Conference Proceedings
·
OSTI ID:21179510
Progress of the APS high heat load X-ray beam position monitor development
Conference
·
Mon Sep 01 00:00:00 EDT 1997
·
OSTI ID:555503