Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Diffraction and coherence in breast ultrasound tomography: a study with a toroidal array

Journal Article · · Medical Physics
OSTI ID:957784

Ultrasound is commonly used as an adjunct to mammography for diagnostic evaluation of suspicions arising from breast cancer screening. As an alternative to conventional sonography that uses hand-held transducers, toroidal array probes that encircle the breast immersed in a water bath have been investigated for ultrasound tomography. In this paper, two sets of experiments performed with a prototype ultrasound scanner on a phantom and a human breast in vivo are used to investigate the effects of diffraction and coherence in ultrasound tomography. Reconstructions obtained with transmission diffraction tomography (TDT) are compared with conventional reflection imaging and computerized ultrasound tomography showing a substantial improvement. The in vivo tests demonstrate that TDT can image the complex boundary of a cancer mass and suggest that it can reveal the anatomy of milk ducts and Cooper's ligaments.

Research Organization:
Los Alamos National Laboratory (LANL)
Sponsoring Organization:
DOE
DOE Contract Number:
AC52-06NA25396
OSTI ID:
957784
Report Number(s):
LA-UR-08-04616; LA-UR-08-4616
Journal Information:
Medical Physics, Journal Name: Medical Physics; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

A novel synthetic aperture technique for breast tomography with toroidal arrays
Conference · Wed Dec 31 23:00:00 EST 2008 · OSTI ID:956545

Breast ultrasound tomography with total-variation regularization
Conference · Wed Dec 31 23:00:00 EST 2008 · OSTI ID:956546

Development of Ultrasound Tomography for Breast Imaging: Technical Assessment
Journal Article · Thu Sep 30 00:00:00 EDT 2004 · Medical Physics · OSTI ID:15016771