Constant depression fan system a novel glovebox ventilation system

Conference ·
OSTI ID:95669
 [1]
  1. AME plc., Aldermaston (United Kingdom)

In a conventional glovebox ventilation system the depression within the glovebox under normal operation is controlled by instrumentation. In the event of a breach the pressure within the box rises to atmospheric pressure, this pressure rise is detected by instrumentation which in turn operates a quick opening damper in a high depression extract to achieve a 1 metre/sec (200 fpm) inflow through the breach, which can take up to 2 seconds to establish. This system, although widely used, suffers from two distinct drawbacks: It takes a finite time to achieve the containment velocity of 1 metre/sec. It relies upon instrumentation to achieve its objectives. A new glovebox ventilation system has been developed by AWE to overcome these drawbacks. This is the Constant Depression Fan System (CDFS) which is based on an extract fan with a flat characteristic. This achieves all the requirements for the ventilation of gloveboxes and has the advantages that: It has only one moving part - the extract fan. It requires NO INSTRUMENTATION to achieve its objectives. It achieves the containment velocity of 1 metre/sec in the shortest possible time - approximately 0.2 seconds - and tests have shown that containment is maintained under breach conditions. Thus the CDFS is SAFER, SIMPLER and MORE RELIABLE.

Research Organization:
USDOE Assistant Secretary for Environment, Safety, and Health, Washington, DC (United States). Office of Environmental Guidance; Nuclear Regulatory Commission, Washington, DC (United States). Office of Nuclear Regulatory Research; International Society of Nuclear Air Treatment Technologies, Inc., Batavia, OH (United States); Harvard Univ., Boston, MA (United States). Harvard Air Cleaning Lab.
OSTI ID:
95669
Report Number(s):
NUREG/CP--0141; CONF-940738--; ON: TI95007828
Country of Publication:
United States
Language:
English

Similar Records

Control of Nuclear Gloveboxes and Enclosures Using the No-Moving-Part Vortex Amplifier (VXA)
Conference · Sun Jul 01 00:00:00 EDT 2007 · OSTI ID:21294792

Mechanistic Analysis of Glovebox Fire Propagation
Conference · Sun Jun 13 00:00:00 EDT 1999 · OSTI ID:762703

On criterions of stability of fans operating in parallel
Conference · Thu Jul 01 00:00:00 EDT 1999 · OSTI ID:20020993