High spatial resolution gamma imaging detector based on a 5" diameter R3292 Hamamatsu PSPMT
Journal Article
·
· IEEE Transactions on Nuclear Science
OSTI ID:954612
High resolution imaging gamma-ray detectors were developed using Hamamatsu's 5" diameter R3292 position sensitive PMT (PSPMT) and a variety of crystal scintillator arrays. Special readout techniques were used to maximize the active imaging area while reducing the number of readout channels. Spatial resolutions approaching 1 mm were obtained in a broad energy range from 20 to 511 keV. Results are also presented of coupling the scintillator arrays to the PMT via imaging light guides consisting of acrylic optical fibers.
- Research Organization:
- Thomas Jefferson Lab National Accelerator Facility
- Sponsoring Organization:
- USDOE
- DOE Contract Number:
- AC05-84ER40150
- OSTI ID:
- 954612
- Report Number(s):
- JLAB-PHY-98-23
- Journal Information:
- IEEE Transactions on Nuclear Science, Journal Name: IEEE Transactions on Nuclear Science; ISSN 0018-9499; ISSN IETNAE
- Country of Publication:
- United States
- Language:
- English
Similar Records
High spatial resolution gamma imaging detector based on a 5 inch diameter R3292 Hamamatsu PSPMT
Large-area PSPMT based gamma-ray imager with edge reclamation
Optimization of a mini-gamma camera based on a 2?2 array of Hamamatsu H8500 PSPMTs
Journal Article
·
Mon Jun 01 00:00:00 EDT 1998
· IEEE Transactions on Nuclear Science
·
OSTI ID:624172
Large-area PSPMT based gamma-ray imager with edge reclamation
Conference
·
Thu Sep 21 00:00:00 EDT 2000
· IEEE Transactions on Nuclear Science
·
OSTI ID:15004016
Optimization of a mini-gamma camera based on a 2?2 array of Hamamatsu H8500 PSPMTs
Journal Article
·
Thu Nov 30 23:00:00 EST 2006
· Nucl.Instr.Meth.A
·
OSTI ID:1097037