Fluorescent Optical Liquid-Level Sensor
An optical method of detecting liquid level is presented that uses fluorescence radiation generated in an impurity-doped glass or plastic slab. In operation, the slab is inserted into the liquid and pump light is coupled into it so that the light is guided by the slab-air interface above the liquid and escapes into the liquid just below its surface. Since the fluorescence is generated only in that section of the slab above the liquid, the fluorescence power will monotonically decrease with increasing liquid level. Thus, a relationship can be established between any signal proportional to it and the liquid level. Because optical fibers link the pump source and the detector of fluorescence radiation to the sensor, no electrical connections are needed in or near the liquid. Their absence vastly decreases the hazard associated with placing a liquid-level sensor in a potentially explosive environment. A laboratory prototype, consisting of a methyl styrene slab doped with an organic dye, has been built and successfully tested in water. Its response to liquid level when pumped by a tunable argon-ion laser at 476, 488, and 496 nm, and by a blue LED, is presented and shown to be consistent with theory. The fluorescence spectra are also presented and discussed.
- Research Organization:
- Sandia National Lab. (SNL-NM), Albuquerque, NM (United States); Sandia National Lab. (SNL-CA), Livermore, CA (United States)
- Sponsoring Organization:
- US Department of Energy (US)
- DOE Contract Number:
- AC04-94AL85000
- OSTI ID:
- 9483
- Report Number(s):
- SAND99-1934J; TRN: AH200124%%282
- Journal Information:
- Optical Engineering, Other Information: Submitted to Optical Engineering; PBD: 26 Jul 1999
- Country of Publication:
- United States
- Language:
- English
Similar Records
Fluorescent Dye-doped Sol-gel Sensor for Highly Sensitive Carbon Dioxide Gas Detection below Atmospheric Concentrations
Diffusion of Hydrophilic to Hydrophobic Forms of Nile Red in Aqueous C12EO10 Gels by Variable Area Fluorescence Correlation Spectroscopy