CeBr3 as a High-Resolution Gamma-Ray Detector
Lanthanum halide (LaBr3:Ce) scintillators have been well-documented as high-resolution gamma-ray detectors that are operated at room temperature. These scintillators have better resolution (<3% at 662 keV) relative to sodium iodide (NaI(Tl)) scintillators (7% at 662 keV), but the naturally occurring radioactive isotope 138La causes self-activity in the crystal that occludes portions of the gamma-ray spectrum. This selfactivity limits the use of LaBr3:Ce in high-sensitivity applications. Cerium, the dopant in the LaBr3:Ce matrix possesses useful scintillation properties, and its selfactivity is on the order of 3750 times less than La; however, Ce has not been fully characterized as the chief component in a scintillation detector. This work investigated Ce as the key scintillation matrix component in a scintillation detector with the hypothesis that CeBr3 promises energy resolution comparable or superior to LaBr3:Ce. The researchers involved with this work believe that CeBr3 may be the answer to obtaining high-temperature, high-resolution spectra with greater sensitivity than LaBr3:Ce.