Focused two-dimensional antiscatter grid for mammography.
Conference
·
OSTI ID:947519
We are developing freestanding high-aspect-ratio, focused, two-dimensional antiscatter grids for mammography using deep x-ray lithography and copper electroforming. The exposure is performed using x-rays from bending magnet beamline 2-BM at the Advanced Photon Source (APS) of Argonne National Laboratory. A 2.8-mm-thick prototype freestanding copper antiscatter grid with 25 {micro}m-wide parallel cell walls and 550 {micro}m periodicity has been fabricated. The progress in developing a dynamic double-exposure technique to create the grid with the cell walls aligned to a point x-ray source of the mammography system is discussed.
- Research Organization:
- Argonne National Laboratory (ANL)
- Sponsoring Organization:
- SC; NIH
- DOE Contract Number:
- AC02-06CH11357
- OSTI ID:
- 947519
- Report Number(s):
- ANL/XFD/CP-106432
- Country of Publication:
- United States
- Language:
- ENGLISH
Similar Records
Effect of scatter and an antiscatter grid on the performance of a slot-scanning digital mammography system
A beamline for micromachining and micro-characterization at the APS
A beamline for micromachining and micro-characterization at the APS
Journal Article
·
Sat Apr 15 00:00:00 EDT 2006
· Medical Physics
·
OSTI ID:20775111
A beamline for micromachining and micro-characterization at the APS
Journal Article
·
Tue Jul 01 00:00:00 EDT 1997
· AIP Conference Proceedings
·
OSTI ID:21179504
A beamline for micromachining and micro-characterization at the APS
Journal Article
·
Tue Jul 01 00:00:00 EDT 1997
· AIP Conference Proceedings
·
OSTI ID:627876