Title: Linearity Testing of Photovoltaic Cells

Conference ·

Photovoltaic devices are rated in terms of their peak power with respect to a specific spectrum, total irradiance, and temperature. To rate photovoltaic devices, a reference detector is required whose response is linear with total irradiance. This paper describes a procedure to determine the linearity of the short-circuit current (I{sub sc}) versus the total irradiance (E{sub tot}) by illuminating a reference cell with two lamps. A device is linear if the current measured with both lamps illuminating the cell is the same as the sum of the currents with each lamp illuminating the cell. The two-lamp method is insensitive to the light spectra or spatial nonuniformity changing with irradiance. The two-lamp method is rapid, easy to implement, and does not require operator intervention to change the irradiances. The presence of room light only limits the lowest irradiance that can be evaluated. Unlike other methods, the two-lamp method does not allow the current to be corrected for nonlinear effects.

Research Organization:
National Renewable Energy Lab. (NREL), Golden, CO (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC36-99-GO10337
OSTI ID:
944450
Resource Relation:
Conference: [Proceedings] 2006 IEEE 4th World Conference on Photovoltaic Energy Conversion (WCPEC-4), 7-12 May 2006, Waikoloa, Hawaii; Related Information: For preprint version see NREL/CP-520-39854
Country of Publication:
United States
Language:
English

Similar Records

Linearity Testing of Photovoltaic Cells: Preprint
Conference · Mon May 01 00:00:00 EDT 2006 · OSTI ID:944450

Linearity Testing of Photovoltaic Cells (Poster)
Conference · Mon May 01 00:00:00 EDT 2006 · OSTI ID:944450

Linearity Testing of Photovoltaic Cells
Journal Article · Sun Jan 01 00:00:00 EST 2006 · Journal of Undergraduate Research · OSTI ID:944450