Laser Tracker Calibration - Testing the Angle Measurement System -

Conference ·
OSTI ID:943475
Physics experiments at the SLAC National Accelerator Laboratory (SLAC) usually require high accuracy positioning, e. g. 100 {micro}m over a distance of 150 m or 25 {micro}m in a 10 x 10 x 3 meter volume. Laser tracker measurement systems have become one of the most important tools for achieving these accuracies when mapping components. The accuracy of these measurements is related to the manufacturing tolerances of various individual components, the resolutions of measurement systems, the overall precision of the assembly, and how well imperfections can be modeled. As with theodolites and total stations, one can remove the effects of most assembly and calibration errors by measuring targets in both direct and reverse positions and computing the mean to obtain the result. However, this approach does not compensate for errors originating from the encoder system. In order to improve and gain a better understanding of laser tracker angle measurement tolerances we extended our laboratory's capabilities with the addition of a horizontal angle calibration test stand. This setup is based on the use of a high precision rotary table providing an angular accuracy of better than 0.2 arcsec. Presently, our setup permits only tests of the horizontal angle measurement system. A test stand for vertical angle calibration is under construction. Distance measurements (LECOCQ & FUSS, 2000) are compared to an interferometer bench for distances of up to 32 m. Together both tests provide a better understanding of the instrument and how it should be operated. The observations also provide a reasonable estimate of covariance information of the measurements according to their actual performance for network adjustments.
Research Organization:
Stanford Linear Accelerator Center (SLAC)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC02-76SF00515
OSTI ID:
943475
Report Number(s):
SLAC-PUB-13476
Country of Publication:
United States
Language:
English

Similar Records

Laser Tracker Test Facility at SLAC - Progress Report
Conference · Thu Feb 21 23:00:00 EST 2008 · OSTI ID:924097

Precision calibration and systematic error reduction in the long trace profiler
Journal Article · Fri Dec 31 23:00:00 EST 1999 · Optical Engineering · OSTI ID:20215068

The Quality Assurance Measures for Laser Tracker
Technical Report · Tue Aug 07 00:00:00 EDT 2018 · OSTI ID:1480954