An automated sampler for the measurement of nonmethane organic compounds.

Journal Article · · Environ. Sci. Technol.
DOI:https://doi.org/10.1021/es001193t· OSTI ID:942914
An automated sampler has been constructed for the unattended collection of whole-air samples in Summa passivated stainless steel canisters. The sampler consists of a Viton diaphragm pump; a differential-pressure flow controller; a 2-position, 3-port valve; a 10-port multiposition valve; and a digital valve sequence programmer that controls the sequence and timing of the electrically actuated valves. All connecting tubing was constructed from Silcosteel tubing. Two configurations of the automated sampler and a passive collection system were evaluated by comparing the combined sampling precisions of the three systems with the analytic precision, derived from replicate analyses of an ambient air sample. The analytic precision was generally <{+-}5%, with higher values observed for analytes with mixing ratios in the tens of pptv. However, analytic precision values for methanol and ethanol were poor, greater than {+-}20%. Values for sampling precision were greater by about a factor of 2 than the analytic precision. The poor results for the light oxygenated hydrocarbons could be caused by sorption of the analytes in the preconcentration and sampling systems and difficulties in peak integration.
Research Organization:
Argonne National Laboratory (ANL)
Sponsoring Organization:
SC
DOE Contract Number:
AC02-06CH11357
OSTI ID:
942914
Report Number(s):
ANL/ER/JA-35901
Journal Information:
Environ. Sci. Technol., Journal Name: Environ. Sci. Technol. Journal Issue: 3 ; Feb. 1, 2001 Vol. 35; ISSN ESTHAG; ISSN 0013-936X
Country of Publication:
United States
Language:
ENGLISH

Similar Records

Construction of an automated, sequential air sampler for remote monitoring of organic atmospheric pollutants
Technical Report · Tue Jul 01 00:00:00 EDT 1986 · OSTI ID:5182038

Short-term diffusive sampler for nitrogen dioxide monitoring in epidemiology
Journal Article · Thu Oct 31 23:00:00 EST 1991 · Journal of the Air and Waste Management Association; (United States) · OSTI ID:5913374

Polyport atmospheric gas sampler
Patent · Tue Apr 11 00:00:00 EDT 1995 · OSTI ID:35050