A LIQUID-FILLED PROPORTIONAL COUNTER
Journal Article
·
· Physical Review Letters
We report properties of single-wire proportional chambers and multiple-wire ionization chambers filled with liquid xenon. Proportional multiplication is seen for anode fields > 10{sup 6} volts/cm (anode-cathode voltage> 1 kV for a 3.5-{micro}-diam anode wire). The efficiency for detecting ionizing radiation is {approx} 100%. The time resolution of the chambers is {+-} 10{sup -7} sec. We attained a spatial resolution of {+-} 15 {micro} with a multiwire ionization chamber, considerably better than that of any other real-time particle detector.
- Research Organization:
- Ernest Orlando Lawrence Berkeley National Laboratory, Berkeley, CA (US)
- Sponsoring Organization:
- Physics Division
- DOE Contract Number:
- AC02-05CH11231
- OSTI ID:
- 942298
- Report Number(s):
- UCRL-20811
- Journal Information:
- Physical Review Letters, Journal Name: Physical Review Letters; ISSN 0031-9007; ISSN PRLTAO
- Country of Publication:
- United States
- Language:
- English